Serendipitous detection of Hemoglobin G-Ferrara variant with Sysmex DIFF channel
A low DIFF channel fluorescence (Sysmex XE technology) may reveal the presence of a variant hemoglobin. In this patient Hemoglobin G Ferrara was demonstrated. [Display omitted] •Greater area laboratories favor the occurrence of uncommon findings.•Low fluorescence DIFF scattergrams signal may hide th...
Saved in:
Published in: | Clinical biochemistry 2016-01, Vol.49 (1-2), p.192-193 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A low DIFF channel fluorescence (Sysmex XE technology) may reveal the presence of a variant hemoglobin. In this patient Hemoglobin G Ferrara was demonstrated. [Display omitted]
•Greater area laboratories favor the occurrence of uncommon findings.•Low fluorescence DIFF scattergrams signal may hide the presence of Hb G-Ferrara.•Alterations of automated hematology tests can reveal an unsuspected diagnosis. |
---|---|