Loading…

Subcutaneous Implantable Cardioverter Defibrillator Lead Failure due to Twiddler Syndrome

We present a case of Twiddler syndrome in a patient with a subcutaneous implantable cardioverter defibrillator (S‐ICD). The patient presented herself to the outpatient clinic with pain in the left chest. Chest x‐ray confirmed Twiddler syndrome and ICD read‐out revealed lead failure resulting in abse...

Full description

Saved in:
Bibliographic Details
Published in:Pacing and clinical electrophysiology 2015-11, Vol.38 (11), p.1369-1371
Main Authors: KOOIMAN, KIRSTEN M., BROUWER, TOM F., HALM, VOKKO P. VAN, KNOPS, REINOUD E.
Format: Article
Language:English
Subjects:
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c3321-3933fdcc5c157540e613b2231c01e5f199ba71ba37290ece5cc6a9482d032053
cites
container_end_page 1371
container_issue 11
container_start_page 1369
container_title Pacing and clinical electrophysiology
container_volume 38
creator KOOIMAN, KIRSTEN M.
BROUWER, TOM F.
HALM, VOKKO P. VAN
KNOPS, REINOUD E.
description We present a case of Twiddler syndrome in a patient with a subcutaneous implantable cardioverter defibrillator (S‐ICD). The patient presented herself to the outpatient clinic with pain in the left chest. Chest x‐ray confirmed Twiddler syndrome and ICD read‐out revealed lead failure resulting in absent heart rhythm sensing in one vector. The lead and pulse generator were extracted and a new S‐ICD system was reimplanted submuscular underneath the serratus anterior muscle to prevent reoccurrence. Lead investigation revealed an insulation defect caused by excessive mechanical stress.
doi_str_mv 10.1111/pace.12741
format article
fullrecord <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_proquest_miscellaneous_1760915920</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1760915920</sourcerecordid><originalsourceid>FETCH-LOGICAL-c3321-3933fdcc5c157540e613b2231c01e5f199ba71ba37290ece5cc6a9482d032053</originalsourceid><addsrcrecordid>eNo9kMtOwzAQRS0EouWx4QNQlmwCM3YcN0sUykOqgKqVECvLsaeSIWmKkwD9e9IWmI1H8rnW9WHsDOES-7laGUuXyFWCe2yIMoF4hDLbZ0PARMUjMcoG7Khp3gAghUQesgFPBaQKcMheZ11hu9Ysqe6a6KFalWbZmqKkKDfB-fqTQkshuqGFL4IvS9PWIZqQcdGt8WUXKHIdRW0dzb-8c2WPztZLF-qKTtjBwpQNnf6ex2x-O57n9_Hk6e4hv57EVgiOsciEWDhrpUWp-u6Uoig4F2gBSS4wywqjsDBC8QzIkrQ2NVky4g4EBymO2cXu2VWoPzpqWl35xlLfdPsljSqFrNfBoUfPf9GuqMjpVfCVCWv9Z6MHcAd8-ZLW__cIeuNZbzzrrWf9fJ2Pt1ufiXcZ37T0_Z8x4V2nSiipXx7vtJhOnyc397kG8QNZ2n7s</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1760915920</pqid></control><display><type>article</type><title>Subcutaneous Implantable Cardioverter Defibrillator Lead Failure due to Twiddler Syndrome</title><source>Wiley-Blackwell Read &amp; Publish Collection</source><source>SPORTDiscus with Full Text</source><creator>KOOIMAN, KIRSTEN M. ; BROUWER, TOM F. ; HALM, VOKKO P. VAN ; KNOPS, REINOUD E.</creator><creatorcontrib>KOOIMAN, KIRSTEN M. ; BROUWER, TOM F. ; HALM, VOKKO P. VAN ; KNOPS, REINOUD E.</creatorcontrib><description>We present a case of Twiddler syndrome in a patient with a subcutaneous implantable cardioverter defibrillator (S‐ICD). The patient presented herself to the outpatient clinic with pain in the left chest. Chest x‐ray confirmed Twiddler syndrome and ICD read‐out revealed lead failure resulting in absent heart rhythm sensing in one vector. The lead and pulse generator were extracted and a new S‐ICD system was reimplanted submuscular underneath the serratus anterior muscle to prevent reoccurrence. Lead investigation revealed an insulation defect caused by excessive mechanical stress.</description><identifier>ISSN: 0147-8389</identifier><identifier>EISSN: 1540-8159</identifier><identifier>DOI: 10.1111/pace.12741</identifier><identifier>PMID: 26306701</identifier><language>eng</language><publisher>United States: Blackwell Publishing Ltd</publisher><subject>Adult ; complication ; defibrillation - ICD ; Defibrillators, Implantable - adverse effects ; Device Removal ; Female ; Humans ; Prosthesis Failure ; Prosthesis Implantation ; sensing ; Syndrome ; Twiddler</subject><ispartof>Pacing and clinical electrophysiology, 2015-11, Vol.38 (11), p.1369-1371</ispartof><rights>2015 Wiley Periodicals, Inc.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c3321-3933fdcc5c157540e613b2231c01e5f199ba71ba37290ece5cc6a9482d032053</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27923,27924</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/26306701$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>KOOIMAN, KIRSTEN M.</creatorcontrib><creatorcontrib>BROUWER, TOM F.</creatorcontrib><creatorcontrib>HALM, VOKKO P. VAN</creatorcontrib><creatorcontrib>KNOPS, REINOUD E.</creatorcontrib><title>Subcutaneous Implantable Cardioverter Defibrillator Lead Failure due to Twiddler Syndrome</title><title>Pacing and clinical electrophysiology</title><addtitle>Pacing and Clinical Electrophysiology</addtitle><description>We present a case of Twiddler syndrome in a patient with a subcutaneous implantable cardioverter defibrillator (S‐ICD). The patient presented herself to the outpatient clinic with pain in the left chest. Chest x‐ray confirmed Twiddler syndrome and ICD read‐out revealed lead failure resulting in absent heart rhythm sensing in one vector. The lead and pulse generator were extracted and a new S‐ICD system was reimplanted submuscular underneath the serratus anterior muscle to prevent reoccurrence. Lead investigation revealed an insulation defect caused by excessive mechanical stress.</description><subject>Adult</subject><subject>complication</subject><subject>defibrillation - ICD</subject><subject>Defibrillators, Implantable - adverse effects</subject><subject>Device Removal</subject><subject>Female</subject><subject>Humans</subject><subject>Prosthesis Failure</subject><subject>Prosthesis Implantation</subject><subject>sensing</subject><subject>Syndrome</subject><subject>Twiddler</subject><issn>0147-8389</issn><issn>1540-8159</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2015</creationdate><recordtype>article</recordtype><recordid>eNo9kMtOwzAQRS0EouWx4QNQlmwCM3YcN0sUykOqgKqVECvLsaeSIWmKkwD9e9IWmI1H8rnW9WHsDOES-7laGUuXyFWCe2yIMoF4hDLbZ0PARMUjMcoG7Khp3gAghUQesgFPBaQKcMheZ11hu9Ysqe6a6KFalWbZmqKkKDfB-fqTQkshuqGFL4IvS9PWIZqQcdGt8WUXKHIdRW0dzb-8c2WPztZLF-qKTtjBwpQNnf6ex2x-O57n9_Hk6e4hv57EVgiOsciEWDhrpUWp-u6Uoig4F2gBSS4wywqjsDBC8QzIkrQ2NVky4g4EBymO2cXu2VWoPzpqWl35xlLfdPsljSqFrNfBoUfPf9GuqMjpVfCVCWv9Z6MHcAd8-ZLW__cIeuNZbzzrrWf9fJ2Pt1ufiXcZ37T0_Z8x4V2nSiipXx7vtJhOnyc397kG8QNZ2n7s</recordid><startdate>201511</startdate><enddate>201511</enddate><creator>KOOIMAN, KIRSTEN M.</creator><creator>BROUWER, TOM F.</creator><creator>HALM, VOKKO P. VAN</creator><creator>KNOPS, REINOUD E.</creator><general>Blackwell Publishing Ltd</general><scope>BSCLL</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>7X8</scope></search><sort><creationdate>201511</creationdate><title>Subcutaneous Implantable Cardioverter Defibrillator Lead Failure due to Twiddler Syndrome</title><author>KOOIMAN, KIRSTEN M. ; BROUWER, TOM F. ; HALM, VOKKO P. VAN ; KNOPS, REINOUD E.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c3321-3933fdcc5c157540e613b2231c01e5f199ba71ba37290ece5cc6a9482d032053</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2015</creationdate><topic>Adult</topic><topic>complication</topic><topic>defibrillation - ICD</topic><topic>Defibrillators, Implantable - adverse effects</topic><topic>Device Removal</topic><topic>Female</topic><topic>Humans</topic><topic>Prosthesis Failure</topic><topic>Prosthesis Implantation</topic><topic>sensing</topic><topic>Syndrome</topic><topic>Twiddler</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>KOOIMAN, KIRSTEN M.</creatorcontrib><creatorcontrib>BROUWER, TOM F.</creatorcontrib><creatorcontrib>HALM, VOKKO P. VAN</creatorcontrib><creatorcontrib>KNOPS, REINOUD E.</creatorcontrib><collection>Istex</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>MEDLINE - Academic</collection><jtitle>Pacing and clinical electrophysiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>KOOIMAN, KIRSTEN M.</au><au>BROUWER, TOM F.</au><au>HALM, VOKKO P. VAN</au><au>KNOPS, REINOUD E.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Subcutaneous Implantable Cardioverter Defibrillator Lead Failure due to Twiddler Syndrome</atitle><jtitle>Pacing and clinical electrophysiology</jtitle><addtitle>Pacing and Clinical Electrophysiology</addtitle><date>2015-11</date><risdate>2015</risdate><volume>38</volume><issue>11</issue><spage>1369</spage><epage>1371</epage><pages>1369-1371</pages><issn>0147-8389</issn><eissn>1540-8159</eissn><abstract>We present a case of Twiddler syndrome in a patient with a subcutaneous implantable cardioverter defibrillator (S‐ICD). The patient presented herself to the outpatient clinic with pain in the left chest. Chest x‐ray confirmed Twiddler syndrome and ICD read‐out revealed lead failure resulting in absent heart rhythm sensing in one vector. The lead and pulse generator were extracted and a new S‐ICD system was reimplanted submuscular underneath the serratus anterior muscle to prevent reoccurrence. Lead investigation revealed an insulation defect caused by excessive mechanical stress.</abstract><cop>United States</cop><pub>Blackwell Publishing Ltd</pub><pmid>26306701</pmid><doi>10.1111/pace.12741</doi><tpages>3</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0147-8389
ispartof Pacing and clinical electrophysiology, 2015-11, Vol.38 (11), p.1369-1371
issn 0147-8389
1540-8159
language eng
recordid cdi_proquest_miscellaneous_1760915920
source Wiley-Blackwell Read & Publish Collection; SPORTDiscus with Full Text
subjects Adult
complication
defibrillation - ICD
Defibrillators, Implantable - adverse effects
Device Removal
Female
Humans
Prosthesis Failure
Prosthesis Implantation
sensing
Syndrome
Twiddler
title Subcutaneous Implantable Cardioverter Defibrillator Lead Failure due to Twiddler Syndrome
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-12T13%3A50%3A57IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Subcutaneous%20Implantable%20Cardioverter%20Defibrillator%20Lead%20Failure%20due%20to%20Twiddler%20Syndrome&rft.jtitle=Pacing%20and%20clinical%20electrophysiology&rft.au=KOOIMAN,%20KIRSTEN%20M.&rft.date=2015-11&rft.volume=38&rft.issue=11&rft.spage=1369&rft.epage=1371&rft.pages=1369-1371&rft.issn=0147-8389&rft.eissn=1540-8159&rft_id=info:doi/10.1111/pace.12741&rft_dat=%3Cproquest_pubme%3E1760915920%3C/proquest_pubme%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c3321-3933fdcc5c157540e613b2231c01e5f199ba71ba37290ece5cc6a9482d032053%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=1760915920&rft_id=info:pmid/26306701&rfr_iscdi=true