Loading…
Development of a device for the measurement of the thermal expansion of artificial graphite materials at high temperature
A device that simultaneously measures the dimensional changes of artificial graphite due to thermal expansion by two methods, contact and non-contact, has been developed. The contact method uses high-density isotropic graphite as a reference material, measurement jigs and a heater in electric furnac...
Saved in:
Published in: | Carbon (New York) 2015-11, Vol.93, p.1079-1079 |
---|---|
Main Author: | |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A device that simultaneously measures the dimensional changes of artificial graphite due to thermal expansion by two methods, contact and non-contact, has been developed. The contact method uses high-density isotropic graphite as a reference material, measurement jigs and a heater in electric furnace. The latter used a laser micro-gauge. An accurate measurement of the linear thermal expansion and the coefficient of thermal expansion at temperatures as high as 2400°C could be made by the device using these two complementary methods.
[TANSO 2015 (No. 268), 135–7.] |
---|---|
ISSN: | 0008-6223 1873-3891 |
DOI: | 10.1016/j.carbon.2015.07.018 |