Thermo-magnetic behaviour of AFM-MFM cantilevers

Atomic force microscopy (AFM) experiments were performed to study the behaviour of AFM cantilevers under an external magnetic field B and temperature field produced by a coil with an iron core. Four cantilever types were studied. Forces were measured for different B values and at various coil-to-can...

Full description

Saved in:
Bibliographic Details
Published in:Measurement science & technology 2015-08, Vol.26 (8), p.85002-9
Main Authors: Kumar, M, Arinero, R, Bergez, W, Tordjeman, Ph
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Atomic force microscopy (AFM) experiments were performed to study the behaviour of AFM cantilevers under an external magnetic field B and temperature field produced by a coil with an iron core. Four cantilever types were studied. Forces were measured for different B values and at various coil-to-cantilever separation distances. The results were analysed on the basis of a phenomenological model. This model contains the contribution of two terms, one monopole-monopole interaction at short distance, and one apparent paramagnetic interaction in at large distance, which represents the temperature effects. We observe a good agreement between the model and the experimental data.
ISSN:0957-0233
1361-6501
DOI:10.1088/0957-0233/26/8/085002