Quench Simulation for 9.4 T MRI Superconducting Magnet
The 9.4 T MRI superconducting magnet is being fabricated. A quench simulation code based on anisotropic continuum model has been developed for the quench protection design. The quench behavior of the magnet is studied by means of the code. A commercial software based on finite element method is also...
Saved in:
Published in: | Journal of physics. Conference series 2014-01, Vol.507 (3), p.32054-4 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The 9.4 T MRI superconducting magnet is being fabricated. A quench simulation code based on anisotropic continuum model has been developed for the quench protection design. The quench behavior of the magnet is studied by means of the code. A commercial software based on finite element method is also applied to the quench simulation of the magnet. The calculation results of the two numerical methods have been compared and analyzed. |
---|---|
ISSN: | 1742-6588 1742-6596 |
DOI: | 10.1088/1742-6596/507/3/032054 |