SUMMARY We designed and evaluated ASICs (application‐specific integrated circuits) for a human activity monitoring system in order to reduce total power consumption. The ASIC extracts the HR from the measured electrocardiogram (ECG) raw signal. To create an ASIC with ultra‐low power consumption, we...

Full description

Saved in:
Bibliographic Details
Published in:Electronics and communications in Japan 2016-04, Vol.99 (4), p.13-20
Main Authors: MATSUMOTO, YUKI, TANAKA, TOMOYA, SONODA, KOJI, KANDA, KENSUKE, FUJITA, TAKAYUKI, MAENAKA, KAZUSUKE
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:SUMMARY We designed and evaluated ASICs (application‐specific integrated circuits) for a human activity monitoring system in order to reduce total power consumption. The ASIC extracts the HR from the measured electrocardiogram (ECG) raw signal. To create an ASIC with ultra‐low power consumption, we used a novel algorithm based on fuzzy logic that is very simple and requires minimal processing power. The fabricated ASIC shows sufficient functionality and a power level as low as 2 μW. A total of 85% of the power consumption in microprocessing unit (MPU) processing was cut by using the heart rate extraction ASIC.
ISSN:1942-9533
1942-9541
DOI:10.1002/ecj.11778