Ultrahigh-Sensitivity Mediator-Free Biosensor Based on a Microfabricated Microwave Resonator for the Detection of Micromolar Glucose Concentrations

This paper presents a miniaturized microwave-resonator-based ultrahigh sensitivity mediator-free biosensor for determining the level of glucose in deionized-water glucose solutions and human sera using integrated passive device technology on a gallium-arsenide substrate. The proposed glucose biosens...

Full description

Saved in:
Bibliographic Details
Published in:IEEE transactions on microwave theory and techniques 2016-01, Vol.64 (1), p.319-327
Main Authors: Adhikari, Kishor Kumar, Kim, Nam-Young
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:This paper presents a miniaturized microwave-resonator-based ultrahigh sensitivity mediator-free biosensor for determining the level of glucose in deionized-water glucose solutions and human sera using integrated passive device technology on a gallium-arsenide substrate. The proposed glucose biosensor, which consists of cross-coupled stepped-impedance resonators (SIRs), strongly concentrates electromagnetic energy between the coupling regions at a central frequency of 6.53 GHz. The changes in effective permeability (μeff) and permittivity (εeff), which correlate with the variations in the glucose concentration, effectively change the equivalent series inductance and shunt capacitance of the biosensor resonator. This concept was used for the first time to develop an ultrahigh-sensitivity biosensor based on a low -Q microwave resonator. The length of each SIR was meandered width wise, and a meandered-line stub-load was embedded inside the SIR to utilize the equivalent high series inductance and shunt capacitance, respectively. The newly designed present biosensor, which linearly detected glucose level within a wide range of concentration, exhibited an ultrahigh sensitivity (978.7 MHz/mgmL -1 for sera) at least 4.918 times higher than that of previously reported microwave-resonator-based glucose biosensors, a lower detection limit of 0.01928 μM, and a rapid detection time of less than 5 s. Supported by S-parameter-based effective-medium-parameter analysis and a sensitivity enhancement principle, the detection accuracy of the proposed biosensor was increased using its glucose-level-dependent spreading of propagation constant (γ) and self-resonances of impedance (Z).
ISSN:0018-9480
1557-9670
DOI:10.1109/TMTT.2015.2503275