Loading…
Microdevice with integrated multi-enzyme sensors for the measurement of pork freshness
An electrochemical microdevice for measuring pork freshness was developed based on the value of measurement indicators—K, Ki, and H. Four sensors were integrated to measure the sum of the concentrations appearing in the equations that define the K, Ki, and H values. With standard solutions, the corr...
Saved in:
Published in: | Sensors and actuators. B, Chemical Chemical, 2016-11, Vol.235, p.535-540 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | An electrochemical microdevice for measuring pork freshness was developed based on the value of measurement indicators—K, Ki, and H. Four sensors were integrated to measure the sum of the concentrations appearing in the equations that define the K, Ki, and H values. With standard solutions, the correlation coefficients for sensors 1–4 were 0.888, 0.979, 0.816, and 0.828, respectively. Each sensor could complete one measurement in 2min. A good correlation was obtained between the K values measured by the proposed microdevice and by ultra-performance liquid chromatography. Based on the sensor responses obtained with real pork extract samples, the R2 value between the K and Ki values was 0.843, while that between the K and H values was only 0.807 during a 15-day storage period. |
---|---|
ISSN: | 0925-4005 1873-3077 |
DOI: | 10.1016/j.snb.2016.05.074 |