The effect of influence quantities and detector orientation on small-field patient-specific IMRT QA: comparison of measurements with various ionization chambers
Intensity-modulated radiation therapy (IMRT) requires a patient-specific quality assurance (QA) program to validate the treatment plan and a high level of dosimetric accuracy in the treatment delivery. Dosimetric verification generally consists of both absolute- and relative-dose measurements in a p...
Saved in:
Published in: | Radiological physics and technology 2017-06, Vol.10 (2), p.195-203 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
cited_by | |
---|---|
cites | cdi_FETCH-LOGICAL-c348t-727823ee50c78e38deafb8cae3cba418728e24d0c625a6824609c8c8c54ee3d43 |
container_end_page | 203 |
container_issue | 2 |
container_start_page | 195 |
container_title | Radiological physics and technology |
container_volume | 10 |
creator | Godson, Henry Finlay Manickam, Ravikumar Saminathan, Sathiyan Ganesh, Kadirampatti Mani Ponmalar, Retna |
description | Intensity-modulated radiation therapy (IMRT) requires a patient-specific quality assurance (QA) program to validate the treatment plan and a high level of dosimetric accuracy in the treatment delivery. Dosimetric verification generally consists of both absolute- and relative-dose measurements in a phantom using ionization chambers. Measurements were carried out with three different ionization chambers (Scanditronix FC 65G, Exradin A18, and PTW PinPoint 31014) to assess the effects of influence quantities such as the stability, pre- and post-irradiation leakage, stem effect, polarity, and ion recombination on the IMRT point-dose verification with two different orientations. The Exradin A18 and PTW PinPoint ion chambers demonstrated noticeable leakage to magnitudes of 0.6 and 1.2%, whereas negligible leakage was observed with FC 65G ion chamber. Maximum deviations of 0.5 and 0.6% were noticed for the smallest field owing to the ion recombination effect with the PTW PinPoint ion chamber in the parallel and perpendicular orientations, respectively. The calculated total uncertainties of all influence quantities for the FC 65G, A18, and PTW PinPoint ion chambers were 0.5, 0.7, and 1.3%, respectively. The uncertainties determined for each chamber were incorporated into the point-dose measurements of 30 head and neck patient-specific QA plans, and the variation was found to be within ±3%. The magnitude of the leakage in a small-volume ion chamber indicated the significance of incorporating the correction factors in the absolute-dose measurement. A paired
t
test analysis indicated that the influence quantities significantly affect the point-dose measurements in the patient-specific IMRT QA. |
doi_str_mv | 10.1007/s12194-016-0385-y |
format | article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_1845816783</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1845816783</sourcerecordid><originalsourceid>FETCH-LOGICAL-c348t-727823ee50c78e38deafb8cae3cba418728e24d0c625a6824609c8c8c54ee3d43</originalsourceid><addsrcrecordid>eNp1kdtqFTEUhoMotlYfwBsJeOPNaE4zk-1dKR4KFVG21yE7s-JOmUmmSaayfRof1bWZWkSQBHJY3_9nkZ-Q55y95oz1bwoXfKMaxruGSd02hwfklOuuxZPiD-_3Up6QJ6VcM9ZxIcRjciL6DRowfkp-bfdAwXtwlSZPQ_TjAtEBvVlsrKEGKNTGgQ5QEUmZphwgVltDihRnmew4Nj7AONAZb7HWlBlc8MHRy09ft_TL-Vvq0jTbHMpR4-kEtiwZJmQL_RHqnt5iMS2Fomn4uXq7vZ12kMtT8sjbscCzu_WMfHv_bnvxsbn6_OHy4vyqcVLp2vSi10ICtMz1GqQewPqddhak21nFdS80CDUw14nWdlqojm2cxtEqADkoeUZerb5zTjcLlGqmUByMo42ArRmuVat512uJ6Mt_0Ou05IjdGY7_KpHrOVJ8pVxOpWTwZs5hsvlgODPH-Mwan8H4zDE-c0DNizvnZTfBcK_4kxcCYgUKluJ3yH89_V_X3xu5qN0</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1910358171</pqid></control><display><type>article</type><title>The effect of influence quantities and detector orientation on small-field patient-specific IMRT QA: comparison of measurements with various ionization chambers</title><source>Springer Nature</source><creator>Godson, Henry Finlay ; Manickam, Ravikumar ; Saminathan, Sathiyan ; Ganesh, Kadirampatti Mani ; Ponmalar, Retna</creator><creatorcontrib>Godson, Henry Finlay ; Manickam, Ravikumar ; Saminathan, Sathiyan ; Ganesh, Kadirampatti Mani ; Ponmalar, Retna</creatorcontrib><description>Intensity-modulated radiation therapy (IMRT) requires a patient-specific quality assurance (QA) program to validate the treatment plan and a high level of dosimetric accuracy in the treatment delivery. Dosimetric verification generally consists of both absolute- and relative-dose measurements in a phantom using ionization chambers. Measurements were carried out with three different ionization chambers (Scanditronix FC 65G, Exradin A18, and PTW PinPoint 31014) to assess the effects of influence quantities such as the stability, pre- and post-irradiation leakage, stem effect, polarity, and ion recombination on the IMRT point-dose verification with two different orientations. The Exradin A18 and PTW PinPoint ion chambers demonstrated noticeable leakage to magnitudes of 0.6 and 1.2%, whereas negligible leakage was observed with FC 65G ion chamber. Maximum deviations of 0.5 and 0.6% were noticed for the smallest field owing to the ion recombination effect with the PTW PinPoint ion chamber in the parallel and perpendicular orientations, respectively. The calculated total uncertainties of all influence quantities for the FC 65G, A18, and PTW PinPoint ion chambers were 0.5, 0.7, and 1.3%, respectively. The uncertainties determined for each chamber were incorporated into the point-dose measurements of 30 head and neck patient-specific QA plans, and the variation was found to be within ±3%. The magnitude of the leakage in a small-volume ion chamber indicated the significance of incorporating the correction factors in the absolute-dose measurement. A paired
t
test analysis indicated that the influence quantities significantly affect the point-dose measurements in the patient-specific IMRT QA.</description><identifier>ISSN: 1865-0333</identifier><identifier>EISSN: 1865-0341</identifier><identifier>DOI: 10.1007/s12194-016-0385-y</identifier><identifier>PMID: 27910001</identifier><language>eng</language><publisher>Singapore: Springer Singapore</publisher><subject>Humans ; Imaging ; Ion recombination ; Ionization chambers ; Irradiation ; Leakage ; Mathematical analysis ; Medical and Radiation Physics ; Medicine ; Medicine & Public Health ; Nuclear Medicine ; Orientation ; Polarity ; Precision Medicine ; Quality assurance ; Quality Assurance, Health Care ; Radiation therapy ; Radiology ; Radiometry - instrumentation ; Radiotherapy ; Radiotherapy, Intensity-Modulated ; Stability analysis ; Uncertainty</subject><ispartof>Radiological physics and technology, 2017-06, Vol.10 (2), p.195-203</ispartof><rights>Japanese Society of Radiological Technology and Japan Society of Medical Physics 2016</rights><rights>Copyright Springer Science & Business Media 2017</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c348t-727823ee50c78e38deafb8cae3cba418728e24d0c625a6824609c8c8c54ee3d43</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27922,27923</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/27910001$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Godson, Henry Finlay</creatorcontrib><creatorcontrib>Manickam, Ravikumar</creatorcontrib><creatorcontrib>Saminathan, Sathiyan</creatorcontrib><creatorcontrib>Ganesh, Kadirampatti Mani</creatorcontrib><creatorcontrib>Ponmalar, Retna</creatorcontrib><title>The effect of influence quantities and detector orientation on small-field patient-specific IMRT QA: comparison of measurements with various ionization chambers</title><title>Radiological physics and technology</title><addtitle>Radiol Phys Technol</addtitle><addtitle>Radiol Phys Technol</addtitle><description>Intensity-modulated radiation therapy (IMRT) requires a patient-specific quality assurance (QA) program to validate the treatment plan and a high level of dosimetric accuracy in the treatment delivery. Dosimetric verification generally consists of both absolute- and relative-dose measurements in a phantom using ionization chambers. Measurements were carried out with three different ionization chambers (Scanditronix FC 65G, Exradin A18, and PTW PinPoint 31014) to assess the effects of influence quantities such as the stability, pre- and post-irradiation leakage, stem effect, polarity, and ion recombination on the IMRT point-dose verification with two different orientations. The Exradin A18 and PTW PinPoint ion chambers demonstrated noticeable leakage to magnitudes of 0.6 and 1.2%, whereas negligible leakage was observed with FC 65G ion chamber. Maximum deviations of 0.5 and 0.6% were noticed for the smallest field owing to the ion recombination effect with the PTW PinPoint ion chamber in the parallel and perpendicular orientations, respectively. The calculated total uncertainties of all influence quantities for the FC 65G, A18, and PTW PinPoint ion chambers were 0.5, 0.7, and 1.3%, respectively. The uncertainties determined for each chamber were incorporated into the point-dose measurements of 30 head and neck patient-specific QA plans, and the variation was found to be within ±3%. The magnitude of the leakage in a small-volume ion chamber indicated the significance of incorporating the correction factors in the absolute-dose measurement. A paired
t
test analysis indicated that the influence quantities significantly affect the point-dose measurements in the patient-specific IMRT QA.</description><subject>Humans</subject><subject>Imaging</subject><subject>Ion recombination</subject><subject>Ionization chambers</subject><subject>Irradiation</subject><subject>Leakage</subject><subject>Mathematical analysis</subject><subject>Medical and Radiation Physics</subject><subject>Medicine</subject><subject>Medicine & Public Health</subject><subject>Nuclear Medicine</subject><subject>Orientation</subject><subject>Polarity</subject><subject>Precision Medicine</subject><subject>Quality assurance</subject><subject>Quality Assurance, Health Care</subject><subject>Radiation therapy</subject><subject>Radiology</subject><subject>Radiometry - instrumentation</subject><subject>Radiotherapy</subject><subject>Radiotherapy, Intensity-Modulated</subject><subject>Stability analysis</subject><subject>Uncertainty</subject><issn>1865-0333</issn><issn>1865-0341</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2017</creationdate><recordtype>article</recordtype><recordid>eNp1kdtqFTEUhoMotlYfwBsJeOPNaE4zk-1dKR4KFVG21yE7s-JOmUmmSaayfRof1bWZWkSQBHJY3_9nkZ-Q55y95oz1bwoXfKMaxruGSd02hwfklOuuxZPiD-_3Up6QJ6VcM9ZxIcRjciL6DRowfkp-bfdAwXtwlSZPQ_TjAtEBvVlsrKEGKNTGgQ5QEUmZphwgVltDihRnmew4Nj7AONAZb7HWlBlc8MHRy09ft_TL-Vvq0jTbHMpR4-kEtiwZJmQL_RHqnt5iMS2Fomn4uXq7vZ12kMtT8sjbscCzu_WMfHv_bnvxsbn6_OHy4vyqcVLp2vSi10ICtMz1GqQewPqddhak21nFdS80CDUw14nWdlqojm2cxtEqADkoeUZerb5zTjcLlGqmUByMo42ArRmuVat512uJ6Mt_0Ou05IjdGY7_KpHrOVJ8pVxOpWTwZs5hsvlgODPH-Mwan8H4zDE-c0DNizvnZTfBcK_4kxcCYgUKluJ3yH89_V_X3xu5qN0</recordid><startdate>20170601</startdate><enddate>20170601</enddate><creator>Godson, Henry Finlay</creator><creator>Manickam, Ravikumar</creator><creator>Saminathan, Sathiyan</creator><creator>Ganesh, Kadirampatti Mani</creator><creator>Ponmalar, Retna</creator><general>Springer Singapore</general><general>Springer Nature B.V</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20170601</creationdate><title>The effect of influence quantities and detector orientation on small-field patient-specific IMRT QA: comparison of measurements with various ionization chambers</title><author>Godson, Henry Finlay ; Manickam, Ravikumar ; Saminathan, Sathiyan ; Ganesh, Kadirampatti Mani ; Ponmalar, Retna</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c348t-727823ee50c78e38deafb8cae3cba418728e24d0c625a6824609c8c8c54ee3d43</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2017</creationdate><topic>Humans</topic><topic>Imaging</topic><topic>Ion recombination</topic><topic>Ionization chambers</topic><topic>Irradiation</topic><topic>Leakage</topic><topic>Mathematical analysis</topic><topic>Medical and Radiation Physics</topic><topic>Medicine</topic><topic>Medicine & Public Health</topic><topic>Nuclear Medicine</topic><topic>Orientation</topic><topic>Polarity</topic><topic>Precision Medicine</topic><topic>Quality assurance</topic><topic>Quality Assurance, Health Care</topic><topic>Radiation therapy</topic><topic>Radiology</topic><topic>Radiometry - instrumentation</topic><topic>Radiotherapy</topic><topic>Radiotherapy, Intensity-Modulated</topic><topic>Stability analysis</topic><topic>Uncertainty</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Godson, Henry Finlay</creatorcontrib><creatorcontrib>Manickam, Ravikumar</creatorcontrib><creatorcontrib>Saminathan, Sathiyan</creatorcontrib><creatorcontrib>Ganesh, Kadirampatti Mani</creatorcontrib><creatorcontrib>Ponmalar, Retna</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Radiological physics and technology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Godson, Henry Finlay</au><au>Manickam, Ravikumar</au><au>Saminathan, Sathiyan</au><au>Ganesh, Kadirampatti Mani</au><au>Ponmalar, Retna</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>The effect of influence quantities and detector orientation on small-field patient-specific IMRT QA: comparison of measurements with various ionization chambers</atitle><jtitle>Radiological physics and technology</jtitle><stitle>Radiol Phys Technol</stitle><addtitle>Radiol Phys Technol</addtitle><date>2017-06-01</date><risdate>2017</risdate><volume>10</volume><issue>2</issue><spage>195</spage><epage>203</epage><pages>195-203</pages><issn>1865-0333</issn><eissn>1865-0341</eissn><abstract>Intensity-modulated radiation therapy (IMRT) requires a patient-specific quality assurance (QA) program to validate the treatment plan and a high level of dosimetric accuracy in the treatment delivery. Dosimetric verification generally consists of both absolute- and relative-dose measurements in a phantom using ionization chambers. Measurements were carried out with three different ionization chambers (Scanditronix FC 65G, Exradin A18, and PTW PinPoint 31014) to assess the effects of influence quantities such as the stability, pre- and post-irradiation leakage, stem effect, polarity, and ion recombination on the IMRT point-dose verification with two different orientations. The Exradin A18 and PTW PinPoint ion chambers demonstrated noticeable leakage to magnitudes of 0.6 and 1.2%, whereas negligible leakage was observed with FC 65G ion chamber. Maximum deviations of 0.5 and 0.6% were noticed for the smallest field owing to the ion recombination effect with the PTW PinPoint ion chamber in the parallel and perpendicular orientations, respectively. The calculated total uncertainties of all influence quantities for the FC 65G, A18, and PTW PinPoint ion chambers were 0.5, 0.7, and 1.3%, respectively. The uncertainties determined for each chamber were incorporated into the point-dose measurements of 30 head and neck patient-specific QA plans, and the variation was found to be within ±3%. The magnitude of the leakage in a small-volume ion chamber indicated the significance of incorporating the correction factors in the absolute-dose measurement. A paired
t
test analysis indicated that the influence quantities significantly affect the point-dose measurements in the patient-specific IMRT QA.</abstract><cop>Singapore</cop><pub>Springer Singapore</pub><pmid>27910001</pmid><doi>10.1007/s12194-016-0385-y</doi><tpages>9</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1865-0333 |
ispartof | Radiological physics and technology, 2017-06, Vol.10 (2), p.195-203 |
issn | 1865-0333 1865-0341 |
language | eng |
recordid | cdi_proquest_miscellaneous_1845816783 |
source | Springer Nature |
subjects | Humans Imaging Ion recombination Ionization chambers Irradiation Leakage Mathematical analysis Medical and Radiation Physics Medicine Medicine & Public Health Nuclear Medicine Orientation Polarity Precision Medicine Quality assurance Quality Assurance, Health Care Radiation therapy Radiology Radiometry - instrumentation Radiotherapy Radiotherapy, Intensity-Modulated Stability analysis Uncertainty |
title | The effect of influence quantities and detector orientation on small-field patient-specific IMRT QA: comparison of measurements with various ionization chambers |
url | http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-14T11%3A33%3A02IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=The%20effect%20of%20influence%20quantities%20and%20detector%20orientation%20on%20small-field%20patient-specific%20IMRT%20QA:%20comparison%20of%20measurements%20with%20various%20ionization%20chambers&rft.jtitle=Radiological%20physics%20and%20technology&rft.au=Godson,%20Henry%20Finlay&rft.date=2017-06-01&rft.volume=10&rft.issue=2&rft.spage=195&rft.epage=203&rft.pages=195-203&rft.issn=1865-0333&rft.eissn=1865-0341&rft_id=info:doi/10.1007/s12194-016-0385-y&rft_dat=%3Cproquest_cross%3E1845816783%3C/proquest_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c348t-727823ee50c78e38deafb8cae3cba418728e24d0c625a6824609c8c8c54ee3d43%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=1910358171&rft_id=info:pmid/27910001&rfr_iscdi=true |