Validating the BTrackS Balance Plate as a low cost alternative for the measurement of sway-induced center of pressure
Abstract The BTrackS Balance Plate (BBP) is a low-cost force plate that provides objective balance assessment and true portability for the user. Given that this technology is relatively new, the purpose of the present study was to provide the first center of pressure (COP) validation of the BBP. Two...
Saved in:
Published in: | Journal of biomechanics 2016-12, Vol.49 (16), p.4142-4145 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Abstract The BTrackS Balance Plate (BBP) is a low-cost force plate that provides objective balance assessment and true portability for the user. Given that this technology is relatively new, the purpose of the present study was to provide the first center of pressure (COP) validation of the BBP. Two BBP devices (one new and one used) were compared with a laboratory-grade force plate (LFP) during simultaneous collection of COP that was induced by an inverted pendulum device with human-like sway characteristics. The results of this study showed almost perfect agreement between the BBP devices and the LFP (ICC>0.999), as well as a high degree of BBP accuracy ( |
---|---|