Loading…

Wavelength calibration of a high resolution spectrograph with a partially stabilized 15-GHz astrocomb from 550 to 890 nm

A visible astrocomb spanning 555-890 nm has been implemented on the 10-m Southern African Large Telescope, delivering complete calibration of one channel of its high-resolution spectrograph and an accurate determination of its resolving power. A novel co-coupling method allowed simultaneous observat...

Full description

Saved in:
Bibliographic Details
Published in:Optics express 2017-03, Vol.25 (6), p.6450-6460
Main Authors: McCracken, Richard A, Depagne, Éric, Kuhn, Rudolf B, Erasmus, Nicolas, Crause, Lisa A, Reid, Derryck T
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A visible astrocomb spanning 555-890 nm has been implemented on the 10-m Southern African Large Telescope, delivering complete calibration of one channel of its high-resolution spectrograph and an accurate determination of its resolving power. A novel co-coupling method allowed simultaneous observation of on-sky, Th-Ar lamp and astrocomb channels, reducing the wavelength calibration uncertainty by a factor of two compared to that obtained using only Th-Ar lines. The excellent passive stability of the master frequency comb laser enabled broadband astrocomb generation without the need for carrier-envelope offset frequency locking, and an atomically referenced narrow linewidth diode laser provided an absolute fiducial marker for wavelength calibration. The simple astrocomb architecture enabled routine operation by non-specialists in an actual telescope environment. On-sky spectroscopy results are presented with direct calibration achieved entirely using the astrocomb.
ISSN:1094-4087
1094-4087
DOI:10.1364/OE.25.006450