Effect of interrupted motion on the cyclic fatigue resistance of reciprocating nickel–titanium instruments
Aim To compare the cyclic fatigue resistances of WaveOne Gold (Denstply Sirona, Ballaigues, Switzerland), WaveOne (Dentsply Sirona) and Reciproc (VDW, Munich, Germany) in interrupted reciprocation. Methodology Three groups of 18 instruments each of WaveOne Gold Primary, WaveOne Primary and Reciproc...
Saved in:
Published in: | International endodontic journal 2018-05, Vol.51 (5), p.549-555 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Aim
To compare the cyclic fatigue resistances of WaveOne Gold (Denstply Sirona, Ballaigues, Switzerland), WaveOne (Dentsply Sirona) and Reciproc (VDW, Munich, Germany) in interrupted reciprocation.
Methodology
Three groups of 18 instruments each of WaveOne Gold Primary, WaveOne Primary and Reciproc R25 were subjected to cyclic fatigue tests in reciprocation and interrupted reciprocation. The instruments in group 1 were tested in reciprocation, whereas in groups 2 and 3, reciprocation was interrupted for 1 s in every 5 s and 10 s, respectively. Cyclic fatigue resistance was tested using dynamic test devices in stainless steel, curved canals (60°, r = 5 mm) until fracture, and the time to fracture values were recorded. The lengths of the fractured tips were also recorded, and fractured surfaces were examined with a scanning electron microscope (SEM). Data were evaluated by one‐way analysis of variance and Tukey's tests.
Results
The cyclic fatigue resistance of WaveOne Gold Primary was not affected by the interruptions (P > 0.05) and exhibited significantly greater cyclic fatigue resistance than WaveOne Primary and Reciproc R25 in all groups (P |
---|---|