POCT PT INR — Is it adequate for patient care? A comparison of the Roche Coaguchek XS vs. Stago Star vs. Siemens BCS in patients routinely seen in an anticoagulation clinic
In this study we examined the difference in patient INR values as measured by the POCT CoaguChek XS device and central laboratory Stago Evolution and Siemens BCS XP analyzers. This study composed of 100 warfarin therapy patients and 20 coagulation normal subjects, showed that the difference between...
Saved in:
Published in: | Clinica chimica acta 2017-09, Vol.472, p.139-145 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | In this study we examined the difference in patient INR values as measured by the POCT CoaguChek XS device and central laboratory Stago Evolution and Siemens BCS XP analyzers.
This study composed of 100 warfarin therapy patients and 20 coagulation normal subjects, showed that the difference between the POCT and clinical laboratory values increased with increasing INR and was exacerbated by the use of different thromboplastin reagents by the POCT and central lab.
The CoaguChek XS and on-site Stago analyzers which used human recombinant (ISI=1.01) and rabbit brain thromboplastin (ISI=1.25), respectively, showed reasonable agreement for INR |
---|---|