Background Late-night salivary cortisol level is one of the first-line tests recommended by the Endocrine Society for the diagnosis of endogenous hypercortisolism. Most routine laboratories measure cortisol levels using immunoassay tests which fail to determine low cortisol levels accurately due to...
Saved in:
Published in: | Journal of endocrinological investigation 2018-03, Vol.41 (3), p.315-323 |
---|---|
Main Authors: | , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
cited_by | cdi_FETCH-LOGICAL-c372t-e60ae5f6f3625dcf70de15a9dde4040f45ac2da55c807f70ab215e0f95a772863 |
---|---|
cites | cdi_FETCH-LOGICAL-c372t-e60ae5f6f3625dcf70de15a9dde4040f45ac2da55c807f70ab215e0f95a772863 |
container_end_page | 323 |
container_issue | 3 |
container_start_page | 315 |
container_title | Journal of endocrinological investigation |
container_volume | 41 |
creator | Mészáros, K. Karvaly, G. Márta, Z. Magda, B. Tőke, J. Szücs, N. Tóth, M. Rácz, K. Patócs, A. |
description | Background
Late-night salivary cortisol level is one of the first-line tests recommended by the Endocrine Society for the diagnosis of endogenous hypercortisolism. Most routine laboratories measure cortisol levels using immunoassay tests which fail to determine low cortisol levels accurately due to the numerous interfering substances. A liquid chromatography–tandem mass spectrometry (LC–MS/MS) method with simple and rapid sample preparation was developed for the simultaneous measurement of cortisol and cortisone and its performance in the diagnosis of endogenous hypercortisolism was evaluated.
Methods
324 late-night saliva samples were analyzed from which 272 samples were derived from patients with a suspected diagnosis of endogenous hypercortisolism. Salivary cortisol levels were assayed using an electrochemiluminescent immunoassay (ECLIA, Cortisol II, Roche), and simultaneous measurement of cortisol and cortisone was performed using an LC–MS/MS method.
Results
A strong correlation between cortisol results measured using ECLIA and LC–MS/MS (
r
2
= 0.892) was demonstrated. Receiver operating characteristics (ROC) analysis showed good diagnostic performance of cortisol and cortisone levels assayed using LC–MS/MS method and for cortisol measured using ECLIA.
Conclusions
Late-night salivary cortisol and cortisone are useful parameters for the diagnosis of hypercortisolism. Using samples obtained from patients where the diagnosis of hypercortisolism is extremely challenging cut-off values for midnight salivary cortisol and cortisone measured by LC–MS/MS method were established. |
doi_str_mv | 10.1007/s40618-017-0743-6 |
format | article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_1930486338</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2007484122</sourcerecordid><originalsourceid>FETCH-LOGICAL-c372t-e60ae5f6f3625dcf70de15a9dde4040f45ac2da55c807f70ab215e0f95a772863</originalsourceid><addsrcrecordid>eNp1kU1uFDEQhVsIRELgAGyQJTZsmvin3fYs0YQ_aSIWgbVVscsTR912Y3cnyo47sOR2nARPJgGExMplv6-eS_Wa5jmjrxml6rh0tGe6pUy1VHWi7R80h0xx2mqh-4d_1QfNk1IuKRVKaPW4OeBaM9np7rD5cRJgG1OZgyUTZp_yCNEiSZ4AiXg93BCHVzikCR0pMIQryDfEpjyHkgYC0d1fIpIRoSwZR4wzWUqI26qTzfrnt--nZ8enZ1WfL5Ij12G-ICWM04C3BhmmsDO_fZgyTpBhDik-bR55GAo-uzuPmi_v3n5ef2g3n95_XL_ZtFYoPrfYU0Dpey96Lp31ijpkElbOYUc76jsJljuQ0mqqqgrnnEmkfiVBKa57cdS82vtOOX1dsMxmDMXiMEDEtBTDVoJ2lRO6oi__QS_TkmOdzvCaSF0p47xSbE_ZnErJ6M2Uw1gXZxg1u-TMPjlTkzO75MxuiBd3zsv5iO53x31UFeB7oFQpbjH_-fr_rr8AYJKnMQ</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2007484122</pqid></control><display><type>article</type><title>Diagnostic performance of a newly developed salivary cortisol and cortisone measurement using an LC–MS/MS method with simple and rapid sample preparation</title><source>Springer Nature</source><creator>Mészáros, K. ; Karvaly, G. ; Márta, Z. ; Magda, B. ; Tőke, J. ; Szücs, N. ; Tóth, M. ; Rácz, K. ; Patócs, A.</creator><creatorcontrib>Mészáros, K. ; Karvaly, G. ; Márta, Z. ; Magda, B. ; Tőke, J. ; Szücs, N. ; Tóth, M. ; Rácz, K. ; Patócs, A.</creatorcontrib><description>Background
Late-night salivary cortisol level is one of the first-line tests recommended by the Endocrine Society for the diagnosis of endogenous hypercortisolism. Most routine laboratories measure cortisol levels using immunoassay tests which fail to determine low cortisol levels accurately due to the numerous interfering substances. A liquid chromatography–tandem mass spectrometry (LC–MS/MS) method with simple and rapid sample preparation was developed for the simultaneous measurement of cortisol and cortisone and its performance in the diagnosis of endogenous hypercortisolism was evaluated.
Methods
324 late-night saliva samples were analyzed from which 272 samples were derived from patients with a suspected diagnosis of endogenous hypercortisolism. Salivary cortisol levels were assayed using an electrochemiluminescent immunoassay (ECLIA, Cortisol II, Roche), and simultaneous measurement of cortisol and cortisone was performed using an LC–MS/MS method.
Results
A strong correlation between cortisol results measured using ECLIA and LC–MS/MS (
r
2
= 0.892) was demonstrated. Receiver operating characteristics (ROC) analysis showed good diagnostic performance of cortisol and cortisone levels assayed using LC–MS/MS method and for cortisol measured using ECLIA.
Conclusions
Late-night salivary cortisol and cortisone are useful parameters for the diagnosis of hypercortisolism. Using samples obtained from patients where the diagnosis of hypercortisolism is extremely challenging cut-off values for midnight salivary cortisol and cortisone measured by LC–MS/MS method were established.</description><identifier>ISSN: 1720-8386</identifier><identifier>ISSN: 0391-4097</identifier><identifier>EISSN: 1720-8386</identifier><identifier>DOI: 10.1007/s40618-017-0743-6</identifier><identifier>PMID: 28815484</identifier><language>eng</language><publisher>Cham: Springer International Publishing</publisher><subject>Cortisol ; Diagnosis ; Endocrinology ; Hormones ; Immunoassay ; Liquid chromatography ; Mass spectroscopy ; Medicine ; Medicine & Public Health ; Metabolic Diseases ; Original Article ; Saliva</subject><ispartof>Journal of endocrinological investigation, 2018-03, Vol.41 (3), p.315-323</ispartof><rights>Italian Society of Endocrinology (SIE) 2017</rights><rights>Copyright Springer Science & Business Media 2018</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c372t-e60ae5f6f3625dcf70de15a9dde4040f45ac2da55c807f70ab215e0f95a772863</citedby><cites>FETCH-LOGICAL-c372t-e60ae5f6f3625dcf70de15a9dde4040f45ac2da55c807f70ab215e0f95a772863</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/28815484$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Mészáros, K.</creatorcontrib><creatorcontrib>Karvaly, G.</creatorcontrib><creatorcontrib>Márta, Z.</creatorcontrib><creatorcontrib>Magda, B.</creatorcontrib><creatorcontrib>Tőke, J.</creatorcontrib><creatorcontrib>Szücs, N.</creatorcontrib><creatorcontrib>Tóth, M.</creatorcontrib><creatorcontrib>Rácz, K.</creatorcontrib><creatorcontrib>Patócs, A.</creatorcontrib><title>Diagnostic performance of a newly developed salivary cortisol and cortisone measurement using an LC–MS/MS method with simple and rapid sample preparation</title><title>Journal of endocrinological investigation</title><addtitle>J Endocrinol Invest</addtitle><addtitle>J Endocrinol Invest</addtitle><description>Background
Late-night salivary cortisol level is one of the first-line tests recommended by the Endocrine Society for the diagnosis of endogenous hypercortisolism. Most routine laboratories measure cortisol levels using immunoassay tests which fail to determine low cortisol levels accurately due to the numerous interfering substances. A liquid chromatography–tandem mass spectrometry (LC–MS/MS) method with simple and rapid sample preparation was developed for the simultaneous measurement of cortisol and cortisone and its performance in the diagnosis of endogenous hypercortisolism was evaluated.
Methods
324 late-night saliva samples were analyzed from which 272 samples were derived from patients with a suspected diagnosis of endogenous hypercortisolism. Salivary cortisol levels were assayed using an electrochemiluminescent immunoassay (ECLIA, Cortisol II, Roche), and simultaneous measurement of cortisol and cortisone was performed using an LC–MS/MS method.
Results
A strong correlation between cortisol results measured using ECLIA and LC–MS/MS (
r
2
= 0.892) was demonstrated. Receiver operating characteristics (ROC) analysis showed good diagnostic performance of cortisol and cortisone levels assayed using LC–MS/MS method and for cortisol measured using ECLIA.
Conclusions
Late-night salivary cortisol and cortisone are useful parameters for the diagnosis of hypercortisolism. Using samples obtained from patients where the diagnosis of hypercortisolism is extremely challenging cut-off values for midnight salivary cortisol and cortisone measured by LC–MS/MS method were established.</description><subject>Cortisol</subject><subject>Diagnosis</subject><subject>Endocrinology</subject><subject>Hormones</subject><subject>Immunoassay</subject><subject>Liquid chromatography</subject><subject>Mass spectroscopy</subject><subject>Medicine</subject><subject>Medicine & Public Health</subject><subject>Metabolic Diseases</subject><subject>Original Article</subject><subject>Saliva</subject><issn>1720-8386</issn><issn>0391-4097</issn><issn>1720-8386</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2018</creationdate><recordtype>article</recordtype><recordid>eNp1kU1uFDEQhVsIRELgAGyQJTZsmvin3fYs0YQ_aSIWgbVVscsTR912Y3cnyo47sOR2nARPJgGExMplv6-eS_Wa5jmjrxml6rh0tGe6pUy1VHWi7R80h0xx2mqh-4d_1QfNk1IuKRVKaPW4OeBaM9np7rD5cRJgG1OZgyUTZp_yCNEiSZ4AiXg93BCHVzikCR0pMIQryDfEpjyHkgYC0d1fIpIRoSwZR4wzWUqI26qTzfrnt--nZ8enZ1WfL5Ij12G-ICWM04C3BhmmsDO_fZgyTpBhDik-bR55GAo-uzuPmi_v3n5ef2g3n95_XL_ZtFYoPrfYU0Dpey96Lp31ijpkElbOYUc76jsJljuQ0mqqqgrnnEmkfiVBKa57cdS82vtOOX1dsMxmDMXiMEDEtBTDVoJ2lRO6oi__QS_TkmOdzvCaSF0p47xSbE_ZnErJ6M2Uw1gXZxg1u-TMPjlTkzO75MxuiBd3zsv5iO53x31UFeB7oFQpbjH_-fr_rr8AYJKnMQ</recordid><startdate>20180301</startdate><enddate>20180301</enddate><creator>Mészáros, K.</creator><creator>Karvaly, G.</creator><creator>Márta, Z.</creator><creator>Magda, B.</creator><creator>Tőke, J.</creator><creator>Szücs, N.</creator><creator>Tóth, M.</creator><creator>Rácz, K.</creator><creator>Patócs, A.</creator><general>Springer International Publishing</general><general>Springer Nature B.V</general><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20180301</creationdate><title>Diagnostic performance of a newly developed salivary cortisol and cortisone measurement using an LC–MS/MS method with simple and rapid sample preparation</title><author>Mészáros, K. ; Karvaly, G. ; Márta, Z. ; Magda, B. ; Tőke, J. ; Szücs, N. ; Tóth, M. ; Rácz, K. ; Patócs, A.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c372t-e60ae5f6f3625dcf70de15a9dde4040f45ac2da55c807f70ab215e0f95a772863</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2018</creationdate><topic>Cortisol</topic><topic>Diagnosis</topic><topic>Endocrinology</topic><topic>Hormones</topic><topic>Immunoassay</topic><topic>Liquid chromatography</topic><topic>Mass spectroscopy</topic><topic>Medicine</topic><topic>Medicine & Public Health</topic><topic>Metabolic Diseases</topic><topic>Original Article</topic><topic>Saliva</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Mészáros, K.</creatorcontrib><creatorcontrib>Karvaly, G.</creatorcontrib><creatorcontrib>Márta, Z.</creatorcontrib><creatorcontrib>Magda, B.</creatorcontrib><creatorcontrib>Tőke, J.</creatorcontrib><creatorcontrib>Szücs, N.</creatorcontrib><creatorcontrib>Tóth, M.</creatorcontrib><creatorcontrib>Rácz, K.</creatorcontrib><creatorcontrib>Patócs, A.</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Journal of endocrinological investigation</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Mészáros, K.</au><au>Karvaly, G.</au><au>Márta, Z.</au><au>Magda, B.</au><au>Tőke, J.</au><au>Szücs, N.</au><au>Tóth, M.</au><au>Rácz, K.</au><au>Patócs, A.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Diagnostic performance of a newly developed salivary cortisol and cortisone measurement using an LC–MS/MS method with simple and rapid sample preparation</atitle><jtitle>Journal of endocrinological investigation</jtitle><stitle>J Endocrinol Invest</stitle><addtitle>J Endocrinol Invest</addtitle><date>2018-03-01</date><risdate>2018</risdate><volume>41</volume><issue>3</issue><spage>315</spage><epage>323</epage><pages>315-323</pages><issn>1720-8386</issn><issn>0391-4097</issn><eissn>1720-8386</eissn><abstract>Background
Late-night salivary cortisol level is one of the first-line tests recommended by the Endocrine Society for the diagnosis of endogenous hypercortisolism. Most routine laboratories measure cortisol levels using immunoassay tests which fail to determine low cortisol levels accurately due to the numerous interfering substances. A liquid chromatography–tandem mass spectrometry (LC–MS/MS) method with simple and rapid sample preparation was developed for the simultaneous measurement of cortisol and cortisone and its performance in the diagnosis of endogenous hypercortisolism was evaluated.
Methods
324 late-night saliva samples were analyzed from which 272 samples were derived from patients with a suspected diagnosis of endogenous hypercortisolism. Salivary cortisol levels were assayed using an electrochemiluminescent immunoassay (ECLIA, Cortisol II, Roche), and simultaneous measurement of cortisol and cortisone was performed using an LC–MS/MS method.
Results
A strong correlation between cortisol results measured using ECLIA and LC–MS/MS (
r
2
= 0.892) was demonstrated. Receiver operating characteristics (ROC) analysis showed good diagnostic performance of cortisol and cortisone levels assayed using LC–MS/MS method and for cortisol measured using ECLIA.
Conclusions
Late-night salivary cortisol and cortisone are useful parameters for the diagnosis of hypercortisolism. Using samples obtained from patients where the diagnosis of hypercortisolism is extremely challenging cut-off values for midnight salivary cortisol and cortisone measured by LC–MS/MS method were established.</abstract><cop>Cham</cop><pub>Springer International Publishing</pub><pmid>28815484</pmid><doi>10.1007/s40618-017-0743-6</doi><tpages>9</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1720-8386 |
ispartof | Journal of endocrinological investigation, 2018-03, Vol.41 (3), p.315-323 |
issn | 1720-8386 0391-4097 1720-8386 |
language | eng |
recordid | cdi_proquest_miscellaneous_1930486338 |
source | Springer Nature |
subjects | Cortisol Diagnosis Endocrinology Hormones Immunoassay Liquid chromatography Mass spectroscopy Medicine Medicine & Public Health Metabolic Diseases Original Article Saliva |
title | Diagnostic performance of a newly developed salivary cortisol and cortisone measurement using an LC–MS/MS method with simple and rapid sample preparation |
url | http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-24T19%3A12%3A49IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Diagnostic%20performance%20of%20a%20newly%20developed%20salivary%20cortisol%20and%20cortisone%20measurement%20using%20an%20LC%E2%80%93MS/MS%20method%20with%20simple%20and%20rapid%20sample%20preparation&rft.jtitle=Journal%20of%20endocrinological%20investigation&rft.au=M%C3%A9sz%C3%A1ros,%20K.&rft.date=2018-03-01&rft.volume=41&rft.issue=3&rft.spage=315&rft.epage=323&rft.pages=315-323&rft.issn=1720-8386&rft.eissn=1720-8386&rft_id=info:doi/10.1007/s40618-017-0743-6&rft_dat=%3Cproquest_cross%3E2007484122%3C/proquest_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c372t-e60ae5f6f3625dcf70de15a9dde4040f45ac2da55c807f70ab215e0f95a772863%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2007484122&rft_id=info:pmid/28815484&rfr_iscdi=true |