How the detector resolution affects the clinical significance of SBRT pre-treatment quality assurance results

•The detector resolution affects the clinical significance of SBRT pre-treatment QA.•DVH evaluation methods offer a more relevant way to evaluate QA results.•The PTW DVH 4D algorithm for DVH metric has been evaluated for the first time. Aim of this work was to study how the detector resolution can a...

Full description

Saved in:
Bibliographic Details
Published in:Physica medica 2018-05, Vol.49, p.129-134
Main Authors: Bruschi, A., Esposito, M., Pini, S., Ghirelli, A., Zatelli, G., Russo, S.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:•The detector resolution affects the clinical significance of SBRT pre-treatment QA.•DVH evaluation methods offer a more relevant way to evaluate QA results.•The PTW DVH 4D algorithm for DVH metric has been evaluated for the first time. Aim of this work was to study how the detector resolution can affect the clinical significance of SBRT pre-treatment volumetric modulated arc therapy (VMAT) verification results. Three detectors (PTW OCTAVIUS 4D 729, 1500 and 100 SRS) used in five configurations with different resolution were compared: 729, 729 merged, 1500, 1500 merged and 1000 SRS. Absolute local gamma passing rates of 3D pre-treatment quality assurance (QA) were evaluated for 150 dose distributions in 30 plans. Five different kinds of error were introduced in order to establish the detection sensitivity of the three devices. Percentage dosimetric differences were evaluated between planned dosevolume histogram (DVH) and patients’ predicted DVH calculated by PTW DVH 4D® software. Results:The mean gamma passing rates and the standard deviations were 92.4% ± 3.7%, 94.6% ± 1.8%, 95.3% ± 4.2%, 97.4% ± 2.5% and 97.6% ± 1.4 respectively for 729, 729 merged, 1500, 1500 merged and 1000 SRS with 2% local dose/2mm criterion. The same trend was found on the sensitivity analysis: using a tight gamma analysis criterion (2%L/1mm) only the 1000 SRS detected every kind of error, while 729 and 1500 merged detected three and four kinds of error respectively. Regarding dose metrics extracted from DVH curves, D50% was within the tolerance level in more than 90% of cases only for the 1000 SRS. The detector resolution can significantly affect the clinical significance of SBRT pre-treatment verification results. The choice of a detector with resolution suitable to the investigated field size is of main importance to avoid getting false positive.
ISSN:1120-1797
1724-191X
DOI:10.1016/j.ejmp.2017.11.012