Loading…
Performance characteristics of the Beckman Coulter UniCel DxI 800 TSH (3rd IS) assay
Beckman Coulter recently reformulated their commercial TSH assay with primary calibration to the World Health Organization 3rd TSH international standard. An extensive evaluation of the performance characteristics for this assay was completed. Intra-day and inter-day precision was evaluated using 3...
Saved in:
Published in: | Clinica chimica acta 2018-03, Vol.478, p.90-100 |
---|---|
Main Authors: | , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Beckman Coulter recently reformulated their commercial TSH assay with primary calibration to the World Health Organization 3rd TSH international standard. An extensive evaluation of the performance characteristics for this assay was completed.
Intra-day and inter-day precision was evaluated using 3 concentrations of commercial quality control material. Linearity, reportable range, stability, sensitivity and susceptibility to common inferences were determined using pooled patient specimens. Inter-assay variability was assessed across 5 different platforms (n=47 patient specimens).
Intra-day and inter-day CVs were |
---|---|
ISSN: | 0009-8981 1873-3492 |
DOI: | 10.1016/j.cca.2017.12.029 |