Loading…

Modular portable unit for thermal conductivity measurements in multiple cryogenic/magnetic field environments

A modular design for a miniature thermal conductivity cell suitable for a dilution refrigerator and other sample-in-vacuum cryogenic environments with different magnet options is described. The sample is mounted once and the contacts remain in place when the unit is repositioned or transported betwe...

Full description

Saved in:
Bibliographic Details
Published in:Review of scientific instruments 2018-01, Vol.89 (1), p.013903-013903
Main Authors: Tanatar, M. A., Bondarenko, V. A., Timmons, E. I., Prozorov, R.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A modular design for a miniature thermal conductivity cell suitable for a dilution refrigerator and other sample-in-vacuum cryogenic environments with different magnet options is described. The sample is mounted once and the contacts remain in place when the unit is repositioned or transported between different magnets and cryostats. This mobility enables comprehensive measurements with access to specific options, such as vector magnet in one lab and ultra-high field in another. This design enables significant expansion of the temperature range by using not only dilution refrigerators but also 3He, 4He cryostats and even ubiquitous Quantum Design Physical Property Measurement System.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.5001708