Leaflet escape in an Edwards TEKNA mitral prosthesis

In almost every type of artificial valve, structural failure has been described. We are reporting on a case of a sudden leaflet escape of an Edwards TEKNA mitral valve prosthesis 12 years after implantation. The patient had a sudden onset of dyspnea and severe pulmonary edema with subsequent cardiog...

Full description

Saved in:
Bibliographic Details
Published in:General thoracic and cardiovascular surgery 2018-12, Vol.66 (12), p.744-747
Main Authors: Higuchi, Shinya, Matsunaga, Iwao, Koga, Kiyokazu, Teshima, Eiichi, Nakashima, Atsuhiro, Tominaga, Ryuji, Ito, Tsuyoshi
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:In almost every type of artificial valve, structural failure has been described. We are reporting on a case of a sudden leaflet escape of an Edwards TEKNA mitral valve prosthesis 12 years after implantation. The patient had a sudden onset of dyspnea and severe pulmonary edema with subsequent cardiogenic shock. An emergency mitral replacement was successfully performed. A multi-detector computed tomography scanning and three-dimensional imaging showed two fragments that had embolized in the terminal aorta and the left common iliac artery. The patient presented visual field abnormality, and postoperative head computed tomography showed watershed cerebral infarction. The escaped leaflet that fractured transversely was removed, following the patient’s recovery, during cardiac surgery.
ISSN:1863-6705
1863-6713
DOI:10.1007/s11748-018-0891-4