Impact of Single-Energy Metal Artifact Reduction on CT image quality in patients with dental hardware
To evaluate whether Canon's Single-Energy Metal Artifact Reduction (SEMAR) algorithm can significantly improve subjective and objective image quality of patients with nonremovable dental hardware undergoing CT imaging of the oral cavity and oropharynx. SEMAR was reconstructed from routine Adapt...
Saved in:
Published in: | Computers in biology and medicine 2018-12, Vol.103, p.161-166 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | To evaluate whether Canon's Single-Energy Metal Artifact Reduction (SEMAR) algorithm can significantly improve subjective and objective image quality of patients with nonremovable dental hardware undergoing CT imaging of the oral cavity and oropharynx.
SEMAR was reconstructed from routine Adaptive Iterative Dose Reduction (AIDR) images in 154 patients (46 females and 108 males; mean age 66.3 ± 10.5 years). Subjective SEMAR and AIDR image quality of the mouth floor, sublingual glands, lymphatic ring and overall impression were evaluated by two independent radiologists on a 6-point scale (1 = very good image quality, 6 = poor image quality) and compared to ratings of an oral and maxillofacial surgeon. Interrater agreement was assessed using the intraclass correlation coefficient (ICC). Objective image analysis was performed by placing regions of interest (ROIs) on the mouth floor and measuring CT attenuation in Hounsfield units (HU) and standard deviation (SD).
SEMAR significantly improved subjective image quality in all evaluated structures for all raters (p |
---|---|