Transcutaneous transmission of photobiomodulation light to the spinal canal of dog as measured from cadaver dogs using a multi-channel intra-spinal probe
The target level photobiomodulation (PBM) irradiances along the thoracic to lumbar segment of the interior spinal canal in six cadaver dogs resulting from surface illumination at 980 nm were measured. Following a lateral hemi-laminectomy, a flexible probe fabricated on a plastic tubular substrate of...
Saved in:
Published in: | Lasers in medical science 2019-10, Vol.34 (8), p.1645-1654 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The target level photobiomodulation (PBM) irradiances along the thoracic to lumbar segment of the interior spinal canal in six cadaver dogs resulting from surface illumination at 980 nm were measured. Following a lateral hemi-laminectomy, a flexible probe fabricated on a plastic tubular substrate of 6.325 mm diameter incorporating nine miniature photodetectors was embedded in the thoracic to lumbar segment of the spinal canal. Intra-spinal irradiances at the nine photodetector sites, spanning an approximate 8 cm length caudal to T13, were measured for various applied powers of continuous wave (CW) surface illumination at 980 nm with a maximal power of 10 W corresponding to a surface irradiance of 3.14 W/cm
2
. The surface illumination conditions differed in skin transmission when the probe was off-contact with tissue and probe-skin contact when the skin was in place. For each condition of surface illumination, the beam was directed to respectively T13 (surface site 1), a spinal column site 4 cm caudal to T13 (surface site 5), and a spinal column site 8 cm caudal to T13 (surface site 9). Off-contact surface irradiation of 3.14 W/cm
2
at surface sites 1, 5, and 9 transmitted respectively 234.0 ± 120.7 μW/cm
2
, 230.7 ± 178.3 μW/cm
2
, and 130.2 ± 169.6 μW/cm
2
to the spinal canal without the skin, and respectively 35.7 ± 33.2 μW/cm
2
, 50.9 ± 75.3 μW/cm
2
, and 15.7 ± 16.3 μW/cm
2
with the skin. Transmission with skin was as low as 12% of the transmission without the skin. On-contact surface irradiation of 3.14 W/cm
2
at surface sites 1, 5, and 9 transmitted respectively 44.6 ± 43.1 μW/cm
2
, 85.4 ± 139.1 μW/cm
2
, and 22.0 ± 23.6 μW/cm
2
to the spinal canal. On-contact application increased transmission by a maximum of 67% comparing to off-contact application. The information gathered highlights the need to clinically consider the impact of skin transmission and on-contact application technique when attempting to treat spinal cord disease with PBM. |
---|---|