An investigation of inter-shock timing and electrode placement for double-sequential defibrillation

Double-Sequential Defibrillation (DSD) is the near-simultaneous use of two defibrillators to treat refractory VF. We hypothesized that (1) risk of DSD-associated defibrillator damage depends on shock vector and (2) the efficacy of DSD depends on inter-shock time. Part 1: risk of defibrillator damage...

Full description

Saved in:
Bibliographic Details
Published in:Resuscitation 2019-07, Vol.140, p.194-200
Main Authors: Taylor, Tyson G., Melnick, Sharon B., Chapman, Fred W., Walcott, Gregory P.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Double-Sequential Defibrillation (DSD) is the near-simultaneous use of two defibrillators to treat refractory VF. We hypothesized that (1) risk of DSD-associated defibrillator damage depends on shock vector and (2) the efficacy of DSD depends on inter-shock time. Part 1: risk of defibrillator damage was assessed in three anaesthetized pigs by applying two sets of defibrillation electrodes in six different configurations (near-orthogonal or near-parallel vectors). Ten 360J shocks were delivered from one set of pads and peak voltage was measured across the second set. Part 2: the dependence of DSD efficacy on inter-shock time was assessed in ten anaesthetized pigs. Electrodes were applied in lateral-lateral (LL) and anterior-posterior positions. Control (LL Stacked Shocks; one vector, two shocks ∼10 s apart) and DSD therapies (Overlapping, 10 ms, 50 ms, 100 ms, 200 ms, 500 ms, 1000 ms apart) were tested in a block randomized design treating electrically-induced VF (n = ∼89 VF episodes/therapy). Shock energies were selected to achieve 25% shock success for a single LL shock. Part 1: peak voltage delivered was 1833 ± 48 V (mean ± 95%CI). Peak voltage exposure was, on average, 10-fold higher for parallel than orthogonal vectors (p 
ISSN:0300-9572
1873-1570
DOI:10.1016/j.resuscitation.2019.04.042