Loading…
Evaluation and calibration of the Model T Coulter Counter used in the population mode
Counts are accurate for 50 and 100 micro m aperture tubes over the entire operable size ranges. Spurious counts, thought to be due to secondary coincidence error, were detected in the lowest 3 sizes when a 200 micro m aperture tube was employed.
Saved in:
Published in: | Analytical chemistry (Washington) 1976-08, Vol.48 (9), p.1415-1418, Article 1415 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Counts are accurate for 50 and 100 micro m aperture tubes over the entire operable size ranges. Spurious counts, thought to be due to secondary coincidence error, were detected in the lowest 3 sizes when a 200 micro m aperture tube was employed. |
---|---|
ISSN: | 0003-2700 1520-6882 |
DOI: | 10.1021/ac50003a036 |