Evaluation and calibration of the Model T Coulter Counter used in the population mode

Counts are accurate for 50 and 100 micro m aperture tubes over the entire operable size ranges. Spurious counts, thought to be due to secondary coincidence error, were detected in the lowest 3 sizes when a 200 micro m aperture tube was employed.

Saved in:
Bibliographic Details
Published in:Analytical chemistry (Washington) 1976-08, Vol.48 (9), p.1415-1418, Article 1415
Main Authors: Sikdar, Subhas K, Randolph, Alan D
Format: Article
Language:English
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Counts are accurate for 50 and 100 micro m aperture tubes over the entire operable size ranges. Spurious counts, thought to be due to secondary coincidence error, were detected in the lowest 3 sizes when a 200 micro m aperture tube was employed.
ISSN:0003-2700
1520-6882