Loading…

Biomechanics of 3-implant-supported and 4-implant-supported mandibular screw-retained prostheses: A 3D finite element analysis study

The number of implants required for the rehabilitation of completely edentulous mandibles has been controversial. The use of a greater number of implants can produce favorable biomechanical outcomes. However, this will lead to high costs and may require complex surgical procedures. Therefore, the mi...

Full description

Saved in:
Bibliographic Details
Published in:The Journal of prosthetic dentistry 2020-07, Vol.124 (1), p.68.e1-68.e10
Main Authors: Elsayyad, Ahmed A., Abbas, Nadia A., AbdelNabi, Nouran M., Osman, Reham B.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The number of implants required for the rehabilitation of completely edentulous mandibles has been controversial. The use of a greater number of implants can produce favorable biomechanical outcomes. However, this will lead to high costs and may require complex surgical procedures. Therefore, the minimum number of implants that can produce desirable outcomes should be used. The purpose of this 3D finite element study was to compare the biomechanics of mandibular 3-implant-supported to 4-implant-supported prostheses. The opposing occlusion was a maxillary complete denture or natural dentition. Two finite element analysis mandibular anatomic models were created. Implants were virtually placed in the mandibular lateral incisor and second premolar region bilaterally in the 4-implant-supported prosthesis model. For the 3-implant-supported model, they were placed in the midline and bilaterally in the second premolar region. Screw-retained polymethyl methacrylate prostheses were designed. Reverse engineering was used to convert standard tessellation language files into computer-aided design solid models. Vertical and oblique loading was applied twice: simulating an opposing maxillary complete denture and a natural dentition. Von Mises stresses and equivalent strains generated in the peri-implant bone, implants’ von Mises stresses and the maximum vertical displacement of the prosthesis were recorded. All recorded outcomes reported higher values for the 3-implant-supported prosthesis compared with the 4-implant-supported models for both applied loads. When opposed by a maxillary complete denture, maximum strain values for the 3-implant-supported (2.3×103 με) and 4-implant-supported (1.6×103 με) models were less than the different threshold limits for the bone resorption reported (3×103, 3.6×103, 6.6×103 με). When opposed by a maxillary natural dentition, maximum strain values for the 3-implant-supported (4.10×103 με) and 4-implant-supported (3.88×103 με) models were less than the highest reported threshold limit for bone resorption (6.6×103 με) in contrast with other reported threshold limits (3×103, 3.6×103 με). In both designs irrespective of the magnitude and direction of loading, the maximum recorded von Mises stresses of the implants (126 MPa) and denture displacement (3.24×102 μm) were less than titanium’s yield strength of (960 to 1180 MPa) and the displacement values (5.2×103 to 8.8×103 μm) reported in the literature. When opposed by a complete denture, rec
ISSN:0022-3913
1097-6841
DOI:10.1016/j.prosdent.2020.01.015