Measurement of complex refractive index with tunable extreme ultraviolet high harmonic source
We report a broadband refractive index measurement method based on a higher harmonic generation tabletop coherent extreme ultraviolet source. We measured the complex refractive index of a sample material by measuring the interference pattern produced by a bare double slit and comparing this with the...
Saved in:
Published in: | Optics express 2020-05, Vol.28 (10), p.15148-15160 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We report a broadband refractive index measurement method based on a higher harmonic generation tabletop coherent extreme ultraviolet source. We measured the complex refractive index of a sample material by measuring the interference pattern produced by a bare double slit and comparing this with the pattern produced by another double slit with one slit covered by the sample material. We validated the method by measuring the complex refractive index of aluminum in the photon energy range of 63-78 eV using a neon gas jet. The measurement system had errors of less than 0.02%. |
---|---|
ISSN: | 1094-4087 1094-4087 |
DOI: | 10.1364/OE.392272 |