Remote robotic percutaneous coronary intervention: An animal feasibility study
Objectives The purpose of this study was to explore the feasibility and safety of robotic PCI performed using an off‐siteremote‐control system in an animal model. Background Access to primary percutaneous coronary intervention (PCI) remains a challenge in acute myocardial infarction management. The...
Saved in:
Published in: | Catheterization and cardiovascular interventions 2021-02, Vol.97 (3), p.E274-E279 |
---|---|
Main Authors: | , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Objectives
The purpose of this study was to explore the feasibility and safety of robotic PCI performed using an off‐siteremote‐control system in an animal model.
Background
Access to primary percutaneous coronary intervention (PCI) remains a challenge in acute myocardial infarction management. The combination of telemedicine and robotic PCI allow the potential delivery of primary PCI to remote locations without the delay of transfer.
Methods
This single‐center prospective pilot preclinical feasibility study compared robotic PCI with remote PCI on swine across three stages (adjacent room, different floor of the same building, two different buildings). Latency up to 1,000 ms was introduced into the operating environment to simulate decreased network quality (blinded to operator). The primary outcome measures were technical success and acute safety. The secondary outcome measures included lesion wiring time, procedural time and qualitative scoring of the PCI experience by the operator.
Results
Across 52 experiments in 15 animals, technical success was 100%. No procedural complications occurred during the study. No significant difference in lesion treatment time was detected between stages (p = .11) and between time per target vessel when latency up to 1,000 ms was introduced (p = .58). Injected delay >250 ms had the greatest impact on procedure perceived lag. Longer procedure time was associated with lower procedure impact score, regardless of injected latency.
Conclusions
Remote robotic PCI was feasible and safe in an animal model. Procedural duration was acceptable and unaffected by network latency. Future studies are needed to determine the safety and feasibility of remote PCI in humans. |
---|---|
ISSN: | 1522-1946 1522-726X |
DOI: | 10.1002/ccd.28978 |