Development and Validation of a Free Carbamazepine Assay on an Automated Chemistry Analyzer
Carbamazepine is an effective drug for treating seizures and trigeminal neuralgia. Therapeutic drug monitoring of free carbamazepine in serum can be useful in situations that drug--protein binding is altered to guide regimen adjustment and to aid in the diagnosis of clinical toxicity. Separation of...
Saved in:
Published in: | The journal of applied laboratory medicine 2020-03, Vol.5 (2), p.357-362 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Carbamazepine is an effective drug for treating seizures and trigeminal neuralgia. Therapeutic drug monitoring of free carbamazepine in serum can be useful in situations that drug--protein binding is altered to guide regimen adjustment and to aid in the diagnosis of clinical toxicity.
Separation of the nonprotein bound carbamazepine was achieved via ultrafiltration through a molecular weight cut-off filter. A method for free carbamazepine measurement was developed on the automated cobas chemistry analyzers (Roche Diagnostics) by modifying the Carbamazepine Gen 4 assay (Roche Diagnostics). Assay performance characteristics were established including precision, accuracy, reportable range, analytical specificity, and stability.
The intra- and inter-assay imprecision was 0%-1.4% and 2.4%-5.1%, respectively. The lower limit of quantitation was 0.3 µg/mL, and the assay was linear up to 10.0 µg/mL. A spike recovery study, using reference standard material, showed recovery was 93.5%--101.3% across the analytical measurement range. Method comparison with a reference laboratory method demonstrated equivalent performance with a slope of 1.01, intercept of 0.09, and correlation coefficient of 0.9948.
This assay provides a simple and accurate method for monitoring free carbamazepine with a fast turnaround time. |
---|---|
ISSN: | 2576-9456 2475-7241 |