Loading…
Quantification of testosterone, androstenedione and 17-hydroxyprogesterone in whole blood collected using Mitra microsampling devices
Background Measurement of testosterone (T), androstenedione (A4) and 17-hydroxyprogesterone (17OHP) usually requires a venous serum sample which may have implications for sample stability or collection. Objective A liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay was developed for sam...
Saved in:
Published in: | Annals of clinical biochemistry 2020-09, Vol.57 (5), p.351-359 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Background
Measurement of testosterone (T), androstenedione (A4) and 17-hydroxyprogesterone (17OHP) usually requires a venous serum sample which may have implications for sample stability or collection.
Objective
A liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay was developed for samples collected using Mitra devices. Analytical validation was completed, and sample comparisons were undertaken to assess Mitra versus venous samples.
Method
Sample was combined with deionized water and internal standard. After mixing, MTBE was added for extraction. The supernatant was transferred to a deep-well plate and dried prior to re-constitution. A HSS T3 column and Waters TQS Micro was used, the detected quantifier transitions were T m/z 289.2 > 96.95, A4 287.2 > 96.95 and 17OHP 331.25 > 96.95.
Results
Mean recovery was 102% for T, 98% for A4 and 97% for 17OHP. Lower limit of quantification was 1 nmol/L for T/A4 and 4 nmol/L for 17OHP. T was linear up to 41.6 nmol/L, A4 41.9 nmol/L and 17OHP 72.6 nmol/L. Ion suppression was |
---|---|
ISSN: | 0004-5632 1758-1001 |