Evaluation of a New Automated Routine Measurement for Serum Adjusted Ionized Calcium (at pH 7.4) in Patients Suspected of Calcium Metabolic Disease

BACKGROUNDTotal calcium is a less accurate test in predicting ionized calcium (Ca2+) in patients suspected of calcium metabolic disease. Nevertheless, total calcium continues to be used as routine measurement instead of adjusted Ca2+ (at pH 7.4). In the current study we evaluate a new multichannel i...

Full description

Saved in:
Bibliographic Details
Published in:The journal of applied laboratory medicine 2020-07, Vol.5 (4), p.704-715
Main Authors: Kur, Dår K, Hillig, Thore, Hansen, Steen I, Goharian, Tina, Witte, Majbritt L, Thode, Jorgen
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:BACKGROUNDTotal calcium is a less accurate test in predicting ionized calcium (Ca2+) in patients suspected of calcium metabolic disease. Nevertheless, total calcium continues to be used as routine measurement instead of adjusted Ca2+ (at pH 7.4). In the current study we evaluate a new multichannel instrument, the ISE Module E1200 for adjusted Ca2+ (at pH 7.4), containing three different ion-selective electrode (ISE) units. METHODSSerum from 1350 patients was compared to the ABL835 flex and KoneLab. Total calcium was also evaluated on the Dimension Vista 1500 system. Correlations between instruments were assessed by Deming regression and degree of agreement by Cohen's kappa (κ). RESULTSAnalytical imprecisions for the three ISE units for adjusted Ca2+ (at pH 7.4) was between 0.36% and 2.52%, and for pH between 0.32% and 3.24%. Results were comparable for each ISE unit (r = 0.797-0.917; all P 
ISSN:2576-9456
2475-7241
DOI:10.1093/jalm/jfaa006