Calibration and applications of polarized neutron thermometry at milli- and microkelvin temperatures
The polarized neutron scattering technique was used for the determination of the nuclear temperature in Cu from 20 mK down to 100 mu K. Using a cooling facility where two adiabatic demagnetization stages work in series the flipping ratio of neutrons scattered from the (200) Bragg-peak was calibrated...
Saved in:
Published in: | Zeitschrift f r Physik B Condensed Matter 1988-06, Vol.71 (2), p.139-148 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The polarized neutron scattering technique was used for the determination of the nuclear temperature in Cu from 20 mK down to 100 mu K. Using a cooling facility where two adiabatic demagnetization stages work in series the flipping ratio of neutrons scattered from the (200) Bragg-peak was calibrated against the nuclear polarization over the whole polarization range. The observed calibration curve deviates considerably from the curve expected for an extinction-free sample. The polarized neutron thermometer is discussed and applied in measurements of the spin--lattice relaxation in Cu. 17 ref.--AA |
---|---|
ISSN: | 0722-3277 1434-6036 |
DOI: | 10.1007/BF01312783 |