Loading…
Vascular sound visualization system is useful for monitoring and surveillance of vascular access
Backgrounds: Vascular access (VA) is a lifeline for maintenance hemodialysis patients. Monitoring and surveillance of VA, therefore, play an important role for maintaining VA patency. Surveillance needs special apparatus to estimate the VA function, while, monitoring including physical examination d...
Saved in:
Published in: | The journal of vascular access 2022-05, Vol.23 (3), p.390-397 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Backgrounds:
Vascular access (VA) is a lifeline for maintenance hemodialysis patients. Monitoring and surveillance of VA, therefore, play an important role for maintaining VA patency. Surveillance needs special apparatus to estimate the VA function, while, monitoring including physical examination does not usually need such devices but highly skill-dependent and qualitative.
Purpose:
We report the clinical utility in monitoring and surveillance of VA using a newly developed electronic stethoscope and vascular sound visualization application.
Methods:
One hundred eight patients participated in the study. The vascular sounds were collected using the electronic stethoscope, converted into digital data, sent wirelessly to a personal computer, and then calculated. The units for the calculated intensity of the vascular sound were decibel [dB]. The intensity, however, was normalized as INDEX values ranging from 0 to 99 for practical use. Correlation between INDEX and the mean flow volume (mFV) and resistance index (RI) of brachial artery measured by Doppler ultrasound sonography were examined for VA surveillance. INDEX was continually measured at each dialysis session in ten patients who underwent the angioplasty for VA monitoring.
Results:
INDEX significantly correlated with the mFV (coefficient of correlation value = 0.68, p |
---|---|
ISSN: | 1129-7298 1724-6032 |
DOI: | 10.1177/1129729821993984 |