Analysis of drug sensitivity of Escherichia Coli O157H7
Sensitive and rapid tests of Escherichia coli drug sensitivity is very important for health of human and animals. An E. coli immunosensor was built based on electrochemical detection and immune detection technologies, through pretreating screen-printed electrodes, and analyzing the optimal reaction...
Saved in:
Published in: | Biomedical microdevices 2021-06, Vol.23 (2), p.29-29, Article 29 |
---|---|
Main Authors: | , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Sensitive and rapid tests of Escherichia coli drug sensitivity is very important for health of human and animals. An
E. coli
immunosensor was built based on electrochemical detection and immune detection technologies, through pretreating screen-printed electrodes, and analyzing the optimal reaction concentration of antigen antibody binding with the AC impedance method. Based on the detection system combining the immunosensor and electrochemical workstation, tests were carried out to measure the accuracy of
E. coli
concentration and drug sensitivity, and error of the detection system was calibrated in accordance with data from the electrochemical workstation.
E. coli
O157:H7 can be detected in the range of 10
3
cfu/ml ~ 10
12
cfu/ml, and the detection error controlled within 5%. Results from the electrochemical workstation and those from the detection device were consistent, and both demonstrated a greater inhibitory effect of antibiotics on
E. coli
than on Bacillus subtilis. The electrochemical detection system is highly efficient and accurate, and could be widely applied to
E. coli
drug sensitivity tests in clinical medicine. |
---|---|