Loading…
Evaluation of Sigma metric approach for monitoring the performance of automated analyzers in hematology unit of Alexandria Main University Hospital
Introduction Sigma metric offers a quantitative framework for evaluating process performance in clinical laboratories. This study aimed to evaluate the analytical performance of automated analyzers in hematology unit of Alexandria Main University Hospital using the sigma metric approach. Materials a...
Saved in:
Published in: | International journal of laboratory hematology 2021-12, Vol.43 (6), p.1388-1393 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Introduction
Sigma metric offers a quantitative framework for evaluating process performance in clinical laboratories. This study aimed to evaluate the analytical performance of automated analyzers in hematology unit of Alexandria Main University Hospital using the sigma metric approach.
Materials and Methods
Quality control data were collected for 6 months, and sigma value was calculated from hematology analyzers SYSMEX (XN 1000, XT 1800i), ADVIA (2120i, 2120), and coagulation analyzers SYSMEX CA 1500 (3610, 6336).
Results
For the normal control level, satisfactory mean sigma value ≥3 was observed for all of the studied parameters by all analyzers. For the high control level, red blood cell count by ADVIA 2120, and hematocrit by ADVIA (2120i and 2120) performed poorly with a mean sigma value |
---|---|
ISSN: | 1751-5521 1751-553X |
DOI: | 10.1111/ijlh.13660 |