Loading…
Safety and efficacy of minimalist transcatheter aortic valve implantation using a new-generation balloon-expandable transcatheter heart valve in bicuspid and tricuspid aortic valves
Background Bicuspid aortic valve may be associated with increased complications during transcatheter aortic valve implantation (TAVI). Aims Compare balloon-expandable transcatheter heart valve (THV) safety and efficacy in severe tricuspid (TAV) and bicuspid (BAV) aortic stenosis. Methods Transfemora...
Saved in:
Published in: | Clinical research in cardiology 2021-12, Vol.110 (12), p.1993-2006 |
---|---|
Main Authors: | , , , , , , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Background
Bicuspid aortic valve may be associated with increased complications during transcatheter aortic valve implantation (TAVI).
Aims
Compare balloon-expandable transcatheter heart valve (THV) safety and efficacy in severe tricuspid (TAV) and bicuspid (BAV) aortic stenosis.
Methods
Transfemoral TAVI was performed in 743 patients (Jan 2014–June 2019) using the SAPIEN 3 THV. Aortic valve morphology was determined using computed tomography. Valve Academic Research Consortium-2 (VARC-2) derived safety and efficacy endpoints at 1 year were evaluated.
Results
BAV patients (
n
= 78), were younger (77 [72, 81] vs. 81 [78, 85] years,
p
|
---|---|
ISSN: | 1861-0684 1861-0692 |
DOI: | 10.1007/s00392-021-01935-7 |