Assessing Phlebotomy Device Preference and Specimen Quality in an Oncology Outpatient Clinic
Oncology patients have frequent venipunctures, which causes scarring, making subsequent draws difficult and painful. Novel blood collection systems may decrease discomfort in patients experiencing repeat blood draws. Oncology outpatients (n = 101; criteria excluded 12) were recruited to determine th...
Saved in:
Published in: | The journal of applied laboratory medicine 2022-03, Vol.7 (2), p.532-540 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
cited_by | cdi_FETCH-LOGICAL-c329t-7293d2fa04c42cf583407348d506a37294b64794f7091167f08ad0b7129793483 |
---|---|
cites | cdi_FETCH-LOGICAL-c329t-7293d2fa04c42cf583407348d506a37294b64794f7091167f08ad0b7129793483 |
container_end_page | 540 |
container_issue | 2 |
container_start_page | 532 |
container_title | The journal of applied laboratory medicine |
container_volume | 7 |
creator | Merrill, VeRonika D Ward, Matthew D Diaz-McNair, Jovita Pickett, Elizabeth A Duh, Show-Hong Christenson, Robert H |
description | Oncology patients have frequent venipunctures, which causes scarring, making subsequent draws difficult and painful. Novel blood collection systems may decrease discomfort in patients experiencing repeat blood draws.
Oncology outpatients (n = 101; criteria excluded 12) were recruited to determine their preference for either of two blood collection systems, the 23-gauge standard BD Vacutainer Push Button Blood Collection Set (Standard Push Button system) or the 25-gauge BD Vacutainer UltraTouch Push Button Blood Collection Set (UltraTouch Push Button system). Subjects received two blinded, randomized blood draws, one with each device and just one device for each arm. Subjects subsequently rated their blinded preference for blood collection system. Specimen quality was assessed for each device with measurements for plasma hemoglobin (Shimadzu UV-1800 spectrophotometer, Shimadzu), lactate dehydrogenase, and potassium (Vitros 4600/5600 analyzer, Ortho Diagnostics).
Preference for the 25-gauge UltraTouch Push Button system over the 23-gauge Standard Push Button system was significant (UltraTouch, n = 51; Standard n = 30; no preference, n = 8; P = 0.0196). Regarding sample quality, the 25-gauge UltraTouch Push Button system had significantly lower plasma hemoglobin (average 5.34 mg/dL) vs the 23-gauge Standard Push Button system (9.37 mg/dL; P |
doi_str_mv | 10.1093/jalm/jfab109 |
format | article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2580939619</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2580939619</sourcerecordid><originalsourceid>FETCH-LOGICAL-c329t-7293d2fa04c42cf583407348d506a37294b64794f7091167f08ad0b7129793483</originalsourceid><addsrcrecordid>eNo9kM1LwzAUwIMobszdPEuOHqzLV5PmOOYnDDZRb0JJ03RmpGltWqH_vRlOT-_rx-O9HwCXGN1iJOlir1y92FeqiNUJmBIm0kQQhk9jngqeSJbyCZiHsEcI4YxwTtE5mFDGKWGSTsHHMgQTgvU7uP10pmj6ph7hnfm22sBtZyrTGR9T5Uv42hpta-Phy6Cc7UdofezDjdeNa3Yj3Ax9q3prfA9XznqrL8BZpVww82OcgfeH-7fVU7LePD6vlutEUyL7eLCkJakUYpoRXaUZZUhQlpUp4orGKSs4E5JVAkmMuahQpkpUCEykkJGjM3D9u7ftmq_BhD6vbdDGOeVNM4ScpFmUJTmWEb35RXXXhBD_y9vO1qobc4zyg9L8oDQ_Ko341XHzUNSm_If_BNIfx7xx2A</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2580939619</pqid></control><display><type>article</type><title>Assessing Phlebotomy Device Preference and Specimen Quality in an Oncology Outpatient Clinic</title><source>Oxford Journals Online</source><creator>Merrill, VeRonika D ; Ward, Matthew D ; Diaz-McNair, Jovita ; Pickett, Elizabeth A ; Duh, Show-Hong ; Christenson, Robert H</creator><creatorcontrib>Merrill, VeRonika D ; Ward, Matthew D ; Diaz-McNair, Jovita ; Pickett, Elizabeth A ; Duh, Show-Hong ; Christenson, Robert H</creatorcontrib><description>Oncology patients have frequent venipunctures, which causes scarring, making subsequent draws difficult and painful. Novel blood collection systems may decrease discomfort in patients experiencing repeat blood draws.
Oncology outpatients (n = 101; criteria excluded 12) were recruited to determine their preference for either of two blood collection systems, the 23-gauge standard BD Vacutainer Push Button Blood Collection Set (Standard Push Button system) or the 25-gauge BD Vacutainer UltraTouch Push Button Blood Collection Set (UltraTouch Push Button system). Subjects received two blinded, randomized blood draws, one with each device and just one device for each arm. Subjects subsequently rated their blinded preference for blood collection system. Specimen quality was assessed for each device with measurements for plasma hemoglobin (Shimadzu UV-1800 spectrophotometer, Shimadzu), lactate dehydrogenase, and potassium (Vitros 4600/5600 analyzer, Ortho Diagnostics).
Preference for the 25-gauge UltraTouch Push Button system over the 23-gauge Standard Push Button system was significant (UltraTouch, n = 51; Standard n = 30; no preference, n = 8; P = 0.0196). Regarding sample quality, the 25-gauge UltraTouch Push Button system had significantly lower plasma hemoglobin (average 5.34 mg/dL) vs the 23-gauge Standard Push Button system (9.37 mg/dL; P < 0.0001); serum lactate dehydrogenase and potassium differences were not statistically significant.
Subjects in an oncology clinic preferred phlebotomy with the 25-gauge UltraTouch Push Button system, and samples using this device had less hemolysis as assessed by plasma hemoglobin.</description><identifier>ISSN: 2576-9456</identifier><identifier>EISSN: 2475-7241</identifier><identifier>DOI: 10.1093/jalm/jfab109</identifier><identifier>PMID: 34632493</identifier><language>eng</language><publisher>England</publisher><subject>Ambulatory Care Facilities ; Blood Specimen Collection ; Hemoglobins - analysis ; Humans ; L-Lactate Dehydrogenase ; Phlebotomy ; Potassium</subject><ispartof>The journal of applied laboratory medicine, 2022-03, Vol.7 (2), p.532-540</ispartof><rights>American Association for Clinical Chemistry 2021.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c329t-7293d2fa04c42cf583407348d506a37294b64794f7091167f08ad0b7129793483</citedby><cites>FETCH-LOGICAL-c329t-7293d2fa04c42cf583407348d506a37294b64794f7091167f08ad0b7129793483</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,778,782,27907,27908</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/34632493$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Merrill, VeRonika D</creatorcontrib><creatorcontrib>Ward, Matthew D</creatorcontrib><creatorcontrib>Diaz-McNair, Jovita</creatorcontrib><creatorcontrib>Pickett, Elizabeth A</creatorcontrib><creatorcontrib>Duh, Show-Hong</creatorcontrib><creatorcontrib>Christenson, Robert H</creatorcontrib><title>Assessing Phlebotomy Device Preference and Specimen Quality in an Oncology Outpatient Clinic</title><title>The journal of applied laboratory medicine</title><addtitle>J Appl Lab Med</addtitle><description>Oncology patients have frequent venipunctures, which causes scarring, making subsequent draws difficult and painful. Novel blood collection systems may decrease discomfort in patients experiencing repeat blood draws.
Oncology outpatients (n = 101; criteria excluded 12) were recruited to determine their preference for either of two blood collection systems, the 23-gauge standard BD Vacutainer Push Button Blood Collection Set (Standard Push Button system) or the 25-gauge BD Vacutainer UltraTouch Push Button Blood Collection Set (UltraTouch Push Button system). Subjects received two blinded, randomized blood draws, one with each device and just one device for each arm. Subjects subsequently rated their blinded preference for blood collection system. Specimen quality was assessed for each device with measurements for plasma hemoglobin (Shimadzu UV-1800 spectrophotometer, Shimadzu), lactate dehydrogenase, and potassium (Vitros 4600/5600 analyzer, Ortho Diagnostics).
Preference for the 25-gauge UltraTouch Push Button system over the 23-gauge Standard Push Button system was significant (UltraTouch, n = 51; Standard n = 30; no preference, n = 8; P = 0.0196). Regarding sample quality, the 25-gauge UltraTouch Push Button system had significantly lower plasma hemoglobin (average 5.34 mg/dL) vs the 23-gauge Standard Push Button system (9.37 mg/dL; P < 0.0001); serum lactate dehydrogenase and potassium differences were not statistically significant.
Subjects in an oncology clinic preferred phlebotomy with the 25-gauge UltraTouch Push Button system, and samples using this device had less hemolysis as assessed by plasma hemoglobin.</description><subject>Ambulatory Care Facilities</subject><subject>Blood Specimen Collection</subject><subject>Hemoglobins - analysis</subject><subject>Humans</subject><subject>L-Lactate Dehydrogenase</subject><subject>Phlebotomy</subject><subject>Potassium</subject><issn>2576-9456</issn><issn>2475-7241</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2022</creationdate><recordtype>article</recordtype><recordid>eNo9kM1LwzAUwIMobszdPEuOHqzLV5PmOOYnDDZRb0JJ03RmpGltWqH_vRlOT-_rx-O9HwCXGN1iJOlir1y92FeqiNUJmBIm0kQQhk9jngqeSJbyCZiHsEcI4YxwTtE5mFDGKWGSTsHHMgQTgvU7uP10pmj6ph7hnfm22sBtZyrTGR9T5Uv42hpta-Phy6Cc7UdofezDjdeNa3Yj3Ax9q3prfA9XznqrL8BZpVww82OcgfeH-7fVU7LePD6vlutEUyL7eLCkJakUYpoRXaUZZUhQlpUp4orGKSs4E5JVAkmMuahQpkpUCEykkJGjM3D9u7ftmq_BhD6vbdDGOeVNM4ScpFmUJTmWEb35RXXXhBD_y9vO1qobc4zyg9L8oDQ_Ko341XHzUNSm_If_BNIfx7xx2A</recordid><startdate>20220302</startdate><enddate>20220302</enddate><creator>Merrill, VeRonika D</creator><creator>Ward, Matthew D</creator><creator>Diaz-McNair, Jovita</creator><creator>Pickett, Elizabeth A</creator><creator>Duh, Show-Hong</creator><creator>Christenson, Robert H</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20220302</creationdate><title>Assessing Phlebotomy Device Preference and Specimen Quality in an Oncology Outpatient Clinic</title><author>Merrill, VeRonika D ; Ward, Matthew D ; Diaz-McNair, Jovita ; Pickett, Elizabeth A ; Duh, Show-Hong ; Christenson, Robert H</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c329t-7293d2fa04c42cf583407348d506a37294b64794f7091167f08ad0b7129793483</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2022</creationdate><topic>Ambulatory Care Facilities</topic><topic>Blood Specimen Collection</topic><topic>Hemoglobins - analysis</topic><topic>Humans</topic><topic>L-Lactate Dehydrogenase</topic><topic>Phlebotomy</topic><topic>Potassium</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Merrill, VeRonika D</creatorcontrib><creatorcontrib>Ward, Matthew D</creatorcontrib><creatorcontrib>Diaz-McNair, Jovita</creatorcontrib><creatorcontrib>Pickett, Elizabeth A</creatorcontrib><creatorcontrib>Duh, Show-Hong</creatorcontrib><creatorcontrib>Christenson, Robert H</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>The journal of applied laboratory medicine</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Merrill, VeRonika D</au><au>Ward, Matthew D</au><au>Diaz-McNair, Jovita</au><au>Pickett, Elizabeth A</au><au>Duh, Show-Hong</au><au>Christenson, Robert H</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Assessing Phlebotomy Device Preference and Specimen Quality in an Oncology Outpatient Clinic</atitle><jtitle>The journal of applied laboratory medicine</jtitle><addtitle>J Appl Lab Med</addtitle><date>2022-03-02</date><risdate>2022</risdate><volume>7</volume><issue>2</issue><spage>532</spage><epage>540</epage><pages>532-540</pages><issn>2576-9456</issn><eissn>2475-7241</eissn><abstract>Oncology patients have frequent venipunctures, which causes scarring, making subsequent draws difficult and painful. Novel blood collection systems may decrease discomfort in patients experiencing repeat blood draws.
Oncology outpatients (n = 101; criteria excluded 12) were recruited to determine their preference for either of two blood collection systems, the 23-gauge standard BD Vacutainer Push Button Blood Collection Set (Standard Push Button system) or the 25-gauge BD Vacutainer UltraTouch Push Button Blood Collection Set (UltraTouch Push Button system). Subjects received two blinded, randomized blood draws, one with each device and just one device for each arm. Subjects subsequently rated their blinded preference for blood collection system. Specimen quality was assessed for each device with measurements for plasma hemoglobin (Shimadzu UV-1800 spectrophotometer, Shimadzu), lactate dehydrogenase, and potassium (Vitros 4600/5600 analyzer, Ortho Diagnostics).
Preference for the 25-gauge UltraTouch Push Button system over the 23-gauge Standard Push Button system was significant (UltraTouch, n = 51; Standard n = 30; no preference, n = 8; P = 0.0196). Regarding sample quality, the 25-gauge UltraTouch Push Button system had significantly lower plasma hemoglobin (average 5.34 mg/dL) vs the 23-gauge Standard Push Button system (9.37 mg/dL; P < 0.0001); serum lactate dehydrogenase and potassium differences were not statistically significant.
Subjects in an oncology clinic preferred phlebotomy with the 25-gauge UltraTouch Push Button system, and samples using this device had less hemolysis as assessed by plasma hemoglobin.</abstract><cop>England</cop><pmid>34632493</pmid><doi>10.1093/jalm/jfab109</doi><tpages>9</tpages><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 2576-9456 |
ispartof | The journal of applied laboratory medicine, 2022-03, Vol.7 (2), p.532-540 |
issn | 2576-9456 2475-7241 |
language | eng |
recordid | cdi_proquest_miscellaneous_2580939619 |
source | Oxford Journals Online |
subjects | Ambulatory Care Facilities Blood Specimen Collection Hemoglobins - analysis Humans L-Lactate Dehydrogenase Phlebotomy Potassium |
title | Assessing Phlebotomy Device Preference and Specimen Quality in an Oncology Outpatient Clinic |
url | http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-16T05%3A23%3A46IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Assessing%20Phlebotomy%20Device%20Preference%20and%20Specimen%20Quality%20in%20an%20Oncology%20Outpatient%20Clinic&rft.jtitle=The%20journal%20of%20applied%20laboratory%20medicine&rft.au=Merrill,%20VeRonika%20D&rft.date=2022-03-02&rft.volume=7&rft.issue=2&rft.spage=532&rft.epage=540&rft.pages=532-540&rft.issn=2576-9456&rft.eissn=2475-7241&rft_id=info:doi/10.1093/jalm/jfab109&rft_dat=%3Cproquest_cross%3E2580939619%3C/proquest_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c329t-7293d2fa04c42cf583407348d506a37294b64794f7091167f08ad0b7129793483%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2580939619&rft_id=info:pmid/34632493&rfr_iscdi=true |