Design and validation of a synchrotron proton beam line for FLASH radiotherapy preclinical research experiments

Purpose The main purpose of this work was to generate and validate the dosimetric accuracy of proton beams of dimensions that are appropriate for in vivo small animal and in vitro ultrahigh dose rate (FLASH) radiotherapy experiments using a synchrotron‐based treatment delivery system. This study was...

Full description

Saved in:
Bibliographic Details
Published in:Medical physics (Lancaster) 2022-01, Vol.49 (1), p.497-509
Main Authors: Titt, Uwe, Yang, Ming, Wang, Xiaochun, Iga, Kiminori, Fredette, Nathaniel, Schueler, Emil, Lin, Steven H, Zhu, X Ron, Sahoo, Narayan, Koong, Albert C, Zhang, Xiaodong, Mohan, Radhe
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c3550-8845aa60ca89a4f74dbb2817a60306018425c8ade7b7fd4a5ed54faccff60873
cites cdi_FETCH-LOGICAL-c3550-8845aa60ca89a4f74dbb2817a60306018425c8ade7b7fd4a5ed54faccff60873
container_end_page 509
container_issue 1
container_start_page 497
container_title Medical physics (Lancaster)
container_volume 49
creator Titt, Uwe
Yang, Ming
Wang, Xiaochun
Iga, Kiminori
Fredette, Nathaniel
Schueler, Emil
Lin, Steven H
Zhu, X Ron
Sahoo, Narayan
Koong, Albert C
Zhang, Xiaodong
Mohan, Radhe
description Purpose The main purpose of this work was to generate and validate the dosimetric accuracy of proton beams of dimensions that are appropriate for in vivo small animal and in vitro ultrahigh dose rate (FLASH) radiotherapy experiments using a synchrotron‐based treatment delivery system. This study was performed to enable future investigations of the relevance of a spread‐out Bragg peak (SOBP) under FLASH conditions. Methods The spill characteristics of the small field fixed horizontal beam line were modified to deliver accelerated protons in times as short as 2 ms and to control the dose delivered. A Gaussian‐like transverse beam profile was transformed into a square uniform one at FLASH dose rates, while avoiding low‐dose regions, a crucial requirement to protect normal tissue during FLASH irradiation. Novel beam‐shaping devices were designed using Monte Carlo techniques to produce up to about 6 cm3 of uniform dose in SOBPs while maximizing the dose rate. These included a scattering foil, a conical flattening filter to maximize the flux of protons into the region of interest, energy filters, range compensators, and collimators. The shapes, sizes, and positions of the components were varied to provide the required field sizes and SOBPs. Results The designed and fabricated devices were used to produce 10‐, 15‐, and 20‐mm diameter, circular field sizes and 10‐, 15‐, and 20‐mm SOBP modulation widths at uniform physical dose rates of up to 375 Gy/s at the center of the SOBP and a minimum dose rate of about 255 Gy/s at the entrance, respectively, in cylindrical volumes. The flatness of lateral dose profiles at the center could be adjusted to within ±1.5% at the center of the SOBP. Assessment of systematic uncertainties, such as impact of misalignments and positioning uncertainties, was performed using simulations, and the results were used to provide appropriate adjustments to ensure high‐accuracy FLASH beam delivery for both in vitro and in vivo preclinical experiments. Conclusions It is feasible to use synchrotron‐generated proton beams of sufficient dimensions for FLASH radiobiology experiments. We expect to use the system we developed to acquire in vitro and in vivo small animal FLASH radiobiology data as a function of dose, dose rate, oxygen content, and linear energy transfer to help us understand the underlying mechanisms of the FLASH phenomenon.
doi_str_mv 10.1002/mp.15370
format article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2600287192</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2600287192</sourcerecordid><originalsourceid>FETCH-LOGICAL-c3550-8845aa60ca89a4f74dbb2817a60306018425c8ade7b7fd4a5ed54faccff60873</originalsourceid><addsrcrecordid>eNp1kEtPwzAQhC0EgvKQ-AXIRy6BTWLH7hEBBaQikOAebZw1NcoLOwX67zGUx4nTaGc_jUbD2GEKJylAdtoOJ6nMFWywSSZUnogMpptsAjAVSSZA7rDdEJ4BoMglbLOdXOh45GrC-gsK7qnj2NX8FRtX4-j6jveWIw-rzix8P_poDFGjVIQtb1xH3Paez-ZnD9fcY-36cUEeh1XkyMS_M9hwT4HQmwWn94G8a6kbwz7bstgEOvjWPfY4u3w8v07md1c352fzxORSQqK1kIgFGNRTFFaJuqoynapo5VBAqkUmjcaaVKVsLVBSLYVFY6wtQKt8jx2vY2PvlyWFsWxdMNQ02FG_DGVWxNm0SqfZH2p8H4InWw6xK_pVmUL5uW7ZDuXXuhE9-k5dVi3Vv-DPnBFI1sCba2j1b1B5e78O_AAQHoQs</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2600287192</pqid></control><display><type>article</type><title>Design and validation of a synchrotron proton beam line for FLASH radiotherapy preclinical research experiments</title><source>Wiley</source><creator>Titt, Uwe ; Yang, Ming ; Wang, Xiaochun ; Iga, Kiminori ; Fredette, Nathaniel ; Schueler, Emil ; Lin, Steven H ; Zhu, X Ron ; Sahoo, Narayan ; Koong, Albert C ; Zhang, Xiaodong ; Mohan, Radhe</creator><creatorcontrib>Titt, Uwe ; Yang, Ming ; Wang, Xiaochun ; Iga, Kiminori ; Fredette, Nathaniel ; Schueler, Emil ; Lin, Steven H ; Zhu, X Ron ; Sahoo, Narayan ; Koong, Albert C ; Zhang, Xiaodong ; Mohan, Radhe</creatorcontrib><description>Purpose The main purpose of this work was to generate and validate the dosimetric accuracy of proton beams of dimensions that are appropriate for in vivo small animal and in vitro ultrahigh dose rate (FLASH) radiotherapy experiments using a synchrotron‐based treatment delivery system. This study was performed to enable future investigations of the relevance of a spread‐out Bragg peak (SOBP) under FLASH conditions. Methods The spill characteristics of the small field fixed horizontal beam line were modified to deliver accelerated protons in times as short as 2 ms and to control the dose delivered. A Gaussian‐like transverse beam profile was transformed into a square uniform one at FLASH dose rates, while avoiding low‐dose regions, a crucial requirement to protect normal tissue during FLASH irradiation. Novel beam‐shaping devices were designed using Monte Carlo techniques to produce up to about 6 cm3 of uniform dose in SOBPs while maximizing the dose rate. These included a scattering foil, a conical flattening filter to maximize the flux of protons into the region of interest, energy filters, range compensators, and collimators. The shapes, sizes, and positions of the components were varied to provide the required field sizes and SOBPs. Results The designed and fabricated devices were used to produce 10‐, 15‐, and 20‐mm diameter, circular field sizes and 10‐, 15‐, and 20‐mm SOBP modulation widths at uniform physical dose rates of up to 375 Gy/s at the center of the SOBP and a minimum dose rate of about 255 Gy/s at the entrance, respectively, in cylindrical volumes. The flatness of lateral dose profiles at the center could be adjusted to within ±1.5% at the center of the SOBP. Assessment of systematic uncertainties, such as impact of misalignments and positioning uncertainties, was performed using simulations, and the results were used to provide appropriate adjustments to ensure high‐accuracy FLASH beam delivery for both in vitro and in vivo preclinical experiments. Conclusions It is feasible to use synchrotron‐generated proton beams of sufficient dimensions for FLASH radiobiology experiments. We expect to use the system we developed to acquire in vitro and in vivo small animal FLASH radiobiology data as a function of dose, dose rate, oxygen content, and linear energy transfer to help us understand the underlying mechanisms of the FLASH phenomenon.</description><identifier>ISSN: 0094-2405</identifier><identifier>EISSN: 2473-4209</identifier><identifier>DOI: 10.1002/mp.15370</identifier><identifier>PMID: 34800037</identifier><language>eng</language><publisher>United States</publisher><subject>Animals ; FLASH therapy ; Monte Carlo Method ; Proton Therapy ; Protons ; Radiotherapy Dosage ; synchrotron beam line ; Synchrotrons</subject><ispartof>Medical physics (Lancaster), 2022-01, Vol.49 (1), p.497-509</ispartof><rights>2021 American Association of Physicists in Medicine</rights><rights>2021 American Association of Physicists in Medicine.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c3550-8845aa60ca89a4f74dbb2817a60306018425c8ade7b7fd4a5ed54faccff60873</citedby><cites>FETCH-LOGICAL-c3550-8845aa60ca89a4f74dbb2817a60306018425c8ade7b7fd4a5ed54faccff60873</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27923,27924</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/34800037$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Titt, Uwe</creatorcontrib><creatorcontrib>Yang, Ming</creatorcontrib><creatorcontrib>Wang, Xiaochun</creatorcontrib><creatorcontrib>Iga, Kiminori</creatorcontrib><creatorcontrib>Fredette, Nathaniel</creatorcontrib><creatorcontrib>Schueler, Emil</creatorcontrib><creatorcontrib>Lin, Steven H</creatorcontrib><creatorcontrib>Zhu, X Ron</creatorcontrib><creatorcontrib>Sahoo, Narayan</creatorcontrib><creatorcontrib>Koong, Albert C</creatorcontrib><creatorcontrib>Zhang, Xiaodong</creatorcontrib><creatorcontrib>Mohan, Radhe</creatorcontrib><title>Design and validation of a synchrotron proton beam line for FLASH radiotherapy preclinical research experiments</title><title>Medical physics (Lancaster)</title><addtitle>Med Phys</addtitle><description>Purpose The main purpose of this work was to generate and validate the dosimetric accuracy of proton beams of dimensions that are appropriate for in vivo small animal and in vitro ultrahigh dose rate (FLASH) radiotherapy experiments using a synchrotron‐based treatment delivery system. This study was performed to enable future investigations of the relevance of a spread‐out Bragg peak (SOBP) under FLASH conditions. Methods The spill characteristics of the small field fixed horizontal beam line were modified to deliver accelerated protons in times as short as 2 ms and to control the dose delivered. A Gaussian‐like transverse beam profile was transformed into a square uniform one at FLASH dose rates, while avoiding low‐dose regions, a crucial requirement to protect normal tissue during FLASH irradiation. Novel beam‐shaping devices were designed using Monte Carlo techniques to produce up to about 6 cm3 of uniform dose in SOBPs while maximizing the dose rate. These included a scattering foil, a conical flattening filter to maximize the flux of protons into the region of interest, energy filters, range compensators, and collimators. The shapes, sizes, and positions of the components were varied to provide the required field sizes and SOBPs. Results The designed and fabricated devices were used to produce 10‐, 15‐, and 20‐mm diameter, circular field sizes and 10‐, 15‐, and 20‐mm SOBP modulation widths at uniform physical dose rates of up to 375 Gy/s at the center of the SOBP and a minimum dose rate of about 255 Gy/s at the entrance, respectively, in cylindrical volumes. The flatness of lateral dose profiles at the center could be adjusted to within ±1.5% at the center of the SOBP. Assessment of systematic uncertainties, such as impact of misalignments and positioning uncertainties, was performed using simulations, and the results were used to provide appropriate adjustments to ensure high‐accuracy FLASH beam delivery for both in vitro and in vivo preclinical experiments. Conclusions It is feasible to use synchrotron‐generated proton beams of sufficient dimensions for FLASH radiobiology experiments. We expect to use the system we developed to acquire in vitro and in vivo small animal FLASH radiobiology data as a function of dose, dose rate, oxygen content, and linear energy transfer to help us understand the underlying mechanisms of the FLASH phenomenon.</description><subject>Animals</subject><subject>FLASH therapy</subject><subject>Monte Carlo Method</subject><subject>Proton Therapy</subject><subject>Protons</subject><subject>Radiotherapy Dosage</subject><subject>synchrotron beam line</subject><subject>Synchrotrons</subject><issn>0094-2405</issn><issn>2473-4209</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2022</creationdate><recordtype>article</recordtype><recordid>eNp1kEtPwzAQhC0EgvKQ-AXIRy6BTWLH7hEBBaQikOAebZw1NcoLOwX67zGUx4nTaGc_jUbD2GEKJylAdtoOJ6nMFWywSSZUnogMpptsAjAVSSZA7rDdEJ4BoMglbLOdXOh45GrC-gsK7qnj2NX8FRtX4-j6jveWIw-rzix8P_poDFGjVIQtb1xH3Paez-ZnD9fcY-36cUEeh1XkyMS_M9hwT4HQmwWn94G8a6kbwz7bstgEOvjWPfY4u3w8v07md1c352fzxORSQqK1kIgFGNRTFFaJuqoynapo5VBAqkUmjcaaVKVsLVBSLYVFY6wtQKt8jx2vY2PvlyWFsWxdMNQ02FG_DGVWxNm0SqfZH2p8H4InWw6xK_pVmUL5uW7ZDuXXuhE9-k5dVi3Vv-DPnBFI1sCba2j1b1B5e78O_AAQHoQs</recordid><startdate>202201</startdate><enddate>202201</enddate><creator>Titt, Uwe</creator><creator>Yang, Ming</creator><creator>Wang, Xiaochun</creator><creator>Iga, Kiminori</creator><creator>Fredette, Nathaniel</creator><creator>Schueler, Emil</creator><creator>Lin, Steven H</creator><creator>Zhu, X Ron</creator><creator>Sahoo, Narayan</creator><creator>Koong, Albert C</creator><creator>Zhang, Xiaodong</creator><creator>Mohan, Radhe</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>202201</creationdate><title>Design and validation of a synchrotron proton beam line for FLASH radiotherapy preclinical research experiments</title><author>Titt, Uwe ; Yang, Ming ; Wang, Xiaochun ; Iga, Kiminori ; Fredette, Nathaniel ; Schueler, Emil ; Lin, Steven H ; Zhu, X Ron ; Sahoo, Narayan ; Koong, Albert C ; Zhang, Xiaodong ; Mohan, Radhe</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c3550-8845aa60ca89a4f74dbb2817a60306018425c8ade7b7fd4a5ed54faccff60873</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2022</creationdate><topic>Animals</topic><topic>FLASH therapy</topic><topic>Monte Carlo Method</topic><topic>Proton Therapy</topic><topic>Protons</topic><topic>Radiotherapy Dosage</topic><topic>synchrotron beam line</topic><topic>Synchrotrons</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Titt, Uwe</creatorcontrib><creatorcontrib>Yang, Ming</creatorcontrib><creatorcontrib>Wang, Xiaochun</creatorcontrib><creatorcontrib>Iga, Kiminori</creatorcontrib><creatorcontrib>Fredette, Nathaniel</creatorcontrib><creatorcontrib>Schueler, Emil</creatorcontrib><creatorcontrib>Lin, Steven H</creatorcontrib><creatorcontrib>Zhu, X Ron</creatorcontrib><creatorcontrib>Sahoo, Narayan</creatorcontrib><creatorcontrib>Koong, Albert C</creatorcontrib><creatorcontrib>Zhang, Xiaodong</creatorcontrib><creatorcontrib>Mohan, Radhe</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Medical physics (Lancaster)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Titt, Uwe</au><au>Yang, Ming</au><au>Wang, Xiaochun</au><au>Iga, Kiminori</au><au>Fredette, Nathaniel</au><au>Schueler, Emil</au><au>Lin, Steven H</au><au>Zhu, X Ron</au><au>Sahoo, Narayan</au><au>Koong, Albert C</au><au>Zhang, Xiaodong</au><au>Mohan, Radhe</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Design and validation of a synchrotron proton beam line for FLASH radiotherapy preclinical research experiments</atitle><jtitle>Medical physics (Lancaster)</jtitle><addtitle>Med Phys</addtitle><date>2022-01</date><risdate>2022</risdate><volume>49</volume><issue>1</issue><spage>497</spage><epage>509</epage><pages>497-509</pages><issn>0094-2405</issn><eissn>2473-4209</eissn><abstract>Purpose The main purpose of this work was to generate and validate the dosimetric accuracy of proton beams of dimensions that are appropriate for in vivo small animal and in vitro ultrahigh dose rate (FLASH) radiotherapy experiments using a synchrotron‐based treatment delivery system. This study was performed to enable future investigations of the relevance of a spread‐out Bragg peak (SOBP) under FLASH conditions. Methods The spill characteristics of the small field fixed horizontal beam line were modified to deliver accelerated protons in times as short as 2 ms and to control the dose delivered. A Gaussian‐like transverse beam profile was transformed into a square uniform one at FLASH dose rates, while avoiding low‐dose regions, a crucial requirement to protect normal tissue during FLASH irradiation. Novel beam‐shaping devices were designed using Monte Carlo techniques to produce up to about 6 cm3 of uniform dose in SOBPs while maximizing the dose rate. These included a scattering foil, a conical flattening filter to maximize the flux of protons into the region of interest, energy filters, range compensators, and collimators. The shapes, sizes, and positions of the components were varied to provide the required field sizes and SOBPs. Results The designed and fabricated devices were used to produce 10‐, 15‐, and 20‐mm diameter, circular field sizes and 10‐, 15‐, and 20‐mm SOBP modulation widths at uniform physical dose rates of up to 375 Gy/s at the center of the SOBP and a minimum dose rate of about 255 Gy/s at the entrance, respectively, in cylindrical volumes. The flatness of lateral dose profiles at the center could be adjusted to within ±1.5% at the center of the SOBP. Assessment of systematic uncertainties, such as impact of misalignments and positioning uncertainties, was performed using simulations, and the results were used to provide appropriate adjustments to ensure high‐accuracy FLASH beam delivery for both in vitro and in vivo preclinical experiments. Conclusions It is feasible to use synchrotron‐generated proton beams of sufficient dimensions for FLASH radiobiology experiments. We expect to use the system we developed to acquire in vitro and in vivo small animal FLASH radiobiology data as a function of dose, dose rate, oxygen content, and linear energy transfer to help us understand the underlying mechanisms of the FLASH phenomenon.</abstract><cop>United States</cop><pmid>34800037</pmid><doi>10.1002/mp.15370</doi><tpages>13</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 0094-2405
ispartof Medical physics (Lancaster), 2022-01, Vol.49 (1), p.497-509
issn 0094-2405
2473-4209
language eng
recordid cdi_proquest_miscellaneous_2600287192
source Wiley
subjects Animals
FLASH therapy
Monte Carlo Method
Proton Therapy
Protons
Radiotherapy Dosage
synchrotron beam line
Synchrotrons
title Design and validation of a synchrotron proton beam line for FLASH radiotherapy preclinical research experiments
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-11T19%3A51%3A04IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Design%20and%20validation%20of%20a%20synchrotron%20proton%20beam%20line%20for%20FLASH%20radiotherapy%20preclinical%20research%20experiments&rft.jtitle=Medical%20physics%20(Lancaster)&rft.au=Titt,%20Uwe&rft.date=2022-01&rft.volume=49&rft.issue=1&rft.spage=497&rft.epage=509&rft.pages=497-509&rft.issn=0094-2405&rft.eissn=2473-4209&rft_id=info:doi/10.1002/mp.15370&rft_dat=%3Cproquest_cross%3E2600287192%3C/proquest_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c3550-8845aa60ca89a4f74dbb2817a60306018425c8ade7b7fd4a5ed54faccff60873%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2600287192&rft_id=info:pmid/34800037&rfr_iscdi=true