Loading…
400 MHz/89 mm actively shielded high resolution NMR magnet
The first actively shielded high resolution NMR magnet was developed and successfully demonstrated. The magnet has excellent field homogeneity in a very large volume and lineshape characteristics exceed the specifications of the standard magnets. Shielding performance is also outstanding, with the 5...
Saved in:
Published in: | IEEE transactions on magnetics 1996-07, Vol.32 (4), p.2647-2650 |
---|---|
Main Authors: | , , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The first actively shielded high resolution NMR magnet was developed and successfully demonstrated. The magnet has excellent field homogeneity in a very large volume and lineshape characteristics exceed the specifications of the standard magnets. Shielding performance is also outstanding, with the 5 G line located only 1 m radially and 1.5 m axially from the magnet center. We are reporting the magnet specifications, details of the design and fabrication, magnet training, field mapping and lineshape results. |
---|---|
ISSN: | 0018-9464 1941-0069 |
DOI: | 10.1109/20.511418 |