Interlaboratory tests of a composite reference sample to calibrate Auger electron spectrometers in the differential mode
A robust reference material has been designed to calibrate the intensity/energy response function of Auger electron spectrometers operated in the differential mode. This material has been tested on six instruments from three manufacturers in five laboratories. The results show that the calibration m...
Saved in:
Published in: | Journal of electron spectroscopy and related phenomena 1993-01, Vol.61 (2), p.173-182 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A robust reference material has been designed to calibrate the intensity/energy response function of Auger electron spectrometers operated in the differential mode. This material has been tested on six instruments from three manufacturers in five laboratories. The results show that the calibration methodology is consistent with previous work for Auger electron spectrometers used in the direct mode and that intensity variations of a factor of two in range, which would occur in quantification without the calibration, may now be reduced to around 5%. |
---|---|
ISSN: | 0368-2048 1873-2526 |