In‐home use of a hybrid closed loop achieves time‐in‐range targets in preschoolers and school children: Results from a randomized, controlled, crossover trial
Aim To obtain additional information on the incremental differences between using a sensor‐augmented pump (SAP) without automated insulin delivery (AID), using it with predictive low‐glucose management (PLGM) or as hybrid closed loop (HCL), in preschool and school children. Methods We conducted a mo...
Saved in:
Published in: | Diabetes, obesity & metabolism obesity & metabolism, 2022-07, Vol.24 (7), p.1319-1327 |
---|---|
Main Authors: | , , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Aim
To obtain additional information on the incremental differences between using a sensor‐augmented pump (SAP) without automated insulin delivery (AID), using it with predictive low‐glucose management (PLGM) or as hybrid closed loop (HCL), in preschool and school children.
Methods
We conducted a monocentric, randomized, controlled, two‐phase crossover study in 38 children aged 2‐6 and 7‐14 years. The primary endpoint was the percentage of time in range (TIR) of 70‐180 mg/dl. Other continuous glucose sensor metrics, HbA1c, patient‐related outcomes (DISABKIDS questionnaire, Fear of Hypoglycaemia Survey) and safety events were also assessed. Results from 2 weeks of SAP, 8 weeks of PLGM and 8 weeks of HCL were compared using a paired t‐test or Wilcoxon signed‐rank test.
Results
Overall, we found a high rate of TIR target (>70%) achievement with HCL in preschool (88%) and school children (50%), with average times in Auto Mode of 93% and 87%, respectively. Preschool children achieved a mean TIR of 73% ± 6% (+8% vs. SAP, +6% vs. PLGM) and school children 69% ± 8% (+15% vs. SAP and + 14% vs. PLGM). Overall, HbA1c improved from 7.4% ± 0.9% to 6.9% ± 0.5% (P = .0002). Diabetes burden and worries and fear of hypoglycaemia remained at low levels, without significant changes versus PLGM. No events of severe hypoglycaemia or diabetic ketoacidosis occurred.
Conclusions
Preschool children profit from AID at least as much as those aged 7 years and older. To ensure safe use and prescribing modalities, regulatory approval is also required for young children. |
---|---|
ISSN: | 1462-8902 1463-1326 |