Frequency of natural antibodies and concordance analysis for anti-TOXOPLASMA IgM tests in Colombian sera of pregnant women
Interpretation of IgM anti-Toxoplasma can be problematic given the phenomena of “natural” IgM. We analyzed 1,503 sera obtained during prenatal care program, and we established natural and false-positive results by doing follow-up. In 101 samples the concordance between enzyme linked immunosorbent as...
Saved in:
Published in: | Diagnostic microbiology and infectious disease 2022-08, Vol.103 (4), p.115733-115733, Article 115733 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Interpretation of IgM anti-Toxoplasma can be problematic given the phenomena of “natural” IgM. We analyzed 1,503 sera obtained during prenatal care program, and we established natural and false-positive results by doing follow-up. In 101 samples the concordance between enzyme linked immunosorbent assay (ELISA) and two semi-automatized systems: electro-chemiluminescence immunoassay (ELECSYS) and vitek immunodiagnostic assay system (VIDAS) was calculated. In 1,503 serum, 71 (4,7%) had ELISA IgG negative and ELISA IgM positive results and in 44 of these had a second sample 4 weeks later. In second samples, 27 (61,3 %) were IgM and IgG negative (false positive result in the first sample) and 13 (29,5%) were ELISA IgM positive and IgG negative (natural IgM). ELISA assay had a poor concordance with enzyme-linked fluorescent immunoassay as well as ELECSYS tests, contrarily, enzyme-linked fluorescent immunoassay and ELECSYS had optimal concordance, with 100 of 101 sera obtaining the same result by both tests. We recommended to use automatized assays to measure IgM anti-Toxoplasma. |
---|---|
ISSN: | 0732-8893 1879-0070 |
DOI: | 10.1016/j.diagmicrobio.2022.115733 |