Loading…
Relapsing low-flow alarms due to suboptimal alignment of the left ventricular assist device inflow cannula
This retrospective study investigated the correlation between the angular position of the left ventricular assist device (LVAD) inflow cannula and relapsing low-flow alarms. Medical charts were reviewed of all patients with HeartMate 3 LVAD support for relapsing low-flow alarms. A standardized proto...
Saved in:
Published in: | European journal of cardio-thoracic surgery 2022-09, Vol.62 (4) |
---|---|
Main Authors: | , , , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | This retrospective study investigated the correlation between the angular position of the left ventricular assist device (LVAD) inflow cannula and relapsing low-flow alarms.
Medical charts were reviewed of all patients with HeartMate 3 LVAD support for relapsing low-flow alarms. A standardized protocol was created to measure the angular position with a contrast-enhanced computed tomography scan. Statistics were done using a gamma frailty model with a constant rate function.
For this analysis, 48 LVAD-supported patients were included. The majority of the patients were male (79%) with a median age of 57 years and a median follow-up of 30 months (interquartile range: 19-41). Low-flow alarm(s) were experienced in 30 (63%) patients. Angulation towards the septal-lateral plane showed a significant increase in low-flow alarms over time with a constant rate function of 0.031 increase in low-flow alarms per month of follow-up per increasing degree of angulation (P = 0.048). When dividing this group using an optimal cut-off point, a significant increase in low-flow alarms was observed when the septal-lateral angulation was 28° or more (P = 0.001). Anterior-posterior and maximal inflow cannula angulation did not show a significant difference.
This study showed an increasing number of low-flow alarms when the degrees of LVAD inflow cannula expand towards the septal-lateral plane. This emphasizes the importance of the LVAD inflow cannula angular position to prevent relapsing low-flow alarms with the risk of diminished quality of life and morbidity. |
---|---|
ISSN: | 1873-734X 1873-734X |
DOI: | 10.1093/ejcts/ezac415 |