SRD1000: a superconductive reference device for thermometry below 1 K
A superconductive reference device (SRD1000), providing 10 reference points in the temperature range 10 mK-1 K with dedicated measurement electronics, has been developed and tested to provide direct traceability to the new provisional low temperature scale (PLTS-2000). We report on the repeatability...
Saved in:
Published in: | Physica. B, Condensed matter Condensed matter, 2003-05, Vol.329-333, p.1562-1563 |
---|---|
Main Author: | |
Format: | Article |
Language: | English |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A superconductive reference device (SRD1000), providing 10 reference points in the temperature range 10 mK-1 K with dedicated measurement electronics, has been developed and tested to provide direct traceability to the new provisional low temperature scale (PLTS-2000). We report on the repeatability of the transition temperatures of samples of IrxRh100-x alloys (with transition temperatures between 20 and 100 mK) and some single crystals (Cd, Zn, AuIn2, AuAl2 and W). |
---|---|
ISSN: | 0921-4526 |
DOI: | 10.1016/S0921-4526(02)02295-0 |