Comparison of Zener voltage standard calibrations at 10 V between the KRISS and the ETL
In order to promote the improvement of the KRISS' 10 V calibration system a comparison of 10 V dc voltage calibrations between the Korea Research Institute of Standards and Science (KRISS, Korea) and the Electrotechnical Laboratory (ETL, Japan) was conducted from Sept.-Oct. 1995. A commercial Z...
Saved in:
Published in: | IEEE transactions on instrumentation and measurement 1997-04, Vol.46 (2), p.314-317 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | In order to promote the improvement of the KRISS' 10 V calibration system a comparison of 10 V dc voltage calibrations between the Korea Research Institute of Standards and Science (KRISS, Korea) and the Electrotechnical Laboratory (ETL, Japan) was conducted from Sept.-Oct. 1995. A commercial Zener voltage standard was carried for the comparison. During the comparison an unexpected discrepancy was found and its origin was experimentally traced to find a correction value. After the correction was made, the result showed the two 10 V calibration systems, both of which are based on 10:1 divider and 1 V Josephson array were in a good agreement within an uncertainty or the comparison which was less than 0.1/spl times/10/sup -6/. Uncertainty of the KRISS system has been improved to be 0.022/spl times/10/sup -6/. |
---|---|