Loading…
Development of an orthogonal-stripe CdZnTe gamma radiation imaging spectrometer
We report performance measurements of a sub-millimeter resolution CdZnTe strip detector developed as a prototype for astronomical instruments operating with good efficiency in the 30-300 keV photon energy range. The prototype is a 1.4 mm thick, 64/spl times/64 contact stripe CdZnTe array of 0.375 mm...
Saved in:
Published in: | IEEE transactions on nuclear science 1996-06, Vol.43 (3), p.1458-1462 |
---|---|
Main Authors: | , , , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We report performance measurements of a sub-millimeter resolution CdZnTe strip detector developed as a prototype for astronomical instruments operating with good efficiency in the 30-300 keV photon energy range. The prototype is a 1.4 mm thick, 64/spl times/64 contact stripe CdZnTe array of 0.375 mm pitch in both dimensions. Pulse height spectra were recorded in orthogonal-stripe coincidence mode which demonstrate room-temperature energy resolution 5:1. Good response is also demonstrated at higher energies using a coplanar grid readout configuration. Spatial resolution capabilities finer than the stripe pitch are demonstrated. We present the image of a /sup 133/Ba source viewed through a collimator slit produced by a 4/spl times/4 stripe detector segment. Charge signals from electron and hole collecting contacts are also discussed. |
---|---|
ISSN: | 0018-9499 1558-1578 |
DOI: | 10.1109/23.507083 |