Loading…

Quantification of cross-vendor variation in ADC measurements in vendor-specific prostate MRI-protocols

•No clinical relevant ADC variations between vendors.•No clinical relevant ADC variations between DWI sequences.•No clinical relevant ADC variations between field strengths. The purpose of this study was to quantify the variability of Apparent Diffusion Coefficient (ADC) and test if there were stati...

Full description

Saved in:
Bibliographic Details
Published in:European journal of radiology 2023-08, Vol.165, p.110942-110942, Article 110942
Main Authors: Møller, Jakob M, Boesen, Lars, Hansen, Adam Espe, Kettles, Karen, Løgager, Vibeke
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:•No clinical relevant ADC variations between vendors.•No clinical relevant ADC variations between DWI sequences.•No clinical relevant ADC variations between field strengths. The purpose of this study was to quantify the variability of Apparent Diffusion Coefficient (ADC) and test if there were statistically significant differences in ADC between MRI systems and sequences. With a two-chamber cylindrical ADC phantom with fixed ADC values (1,000 and 1,600x10−6 mm2/s) a single-shot (ss) Echo Planar Imaging (EPI), a multi-shot EPI, a reduced field of view DWI (zoom) and a Turbo Spin Echo DWI sequence were tested in six MRI systems from three vendors at 1.5 T and 3 T. Technical parameters were according to Prostate Imaging Reporting and Data System Version 2.1. ADC maps were calculated by vendor specific algorithms. Absolute and relative differences in ADC from the phantom-ADC were calculated and differences between sequences were tested. At 3 T absolute differences from phantom given ADC (∼1,000 and ∼ 1,600x10−6 mm2/s) were −83 – 42x10−6 mm2/s (-8.3%-4.2%) and −48 – 15x10−6 mm2/s (-3%-0.9%), respectively and at 1.5 T absolute differences were −81 – 26x10−6 mm2/s (-2.6%-8.1%) and −74 – 67x10−6 mm2/s (-4.6%-4.2%), respectively. Significant statistical differences in ADC measurements were identified between vendors in all sequences except for ssEPI and zoom at 3 T in the 1,600x10−6 mm2/s phantom chamber. Significant differences were also identified between ADC measurements at 1.5 T and 3 T in some of the sequences and vendors, but not all. The variation of ADC between different MRI systems and prostate specific DWI sequences is limited in this phantom study and without apparent clinical relevance. However, prospective multicenter studies of prostate cancer patients are needed for further investigation.
ISSN:0720-048X
1872-7727
DOI:10.1016/j.ejrad.2023.110942