Loading…

Generation of reference electric and magnetic fields for calibration of power-frequency field meters

Improvements in the accuracy of power-frequency electric and magnetic field measurements require adequately calibrated instruments. The systems for the generation of reference electric and magnetic fields set up at the Istituto Elettrotecnico Nazionale 'Galileo Ferraris' (IEN) are describe...

Full description

Saved in:
Bibliographic Details
Published in:IEEE transactions on instrumentation and measurement 1993-04, Vol.42 (2), p.547-552
Main Authors: Bottauscio, O., Crotti, G., D'Emilio, S., Farina, G., Mantini, A.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Improvements in the accuracy of power-frequency electric and magnetic field measurements require adequately calibrated instruments. The systems for the generation of reference electric and magnetic fields set up at the Istituto Elettrotecnico Nazionale 'Galileo Ferraris' (IEN) are described. A description of the calibration procedures is given, together with a detailed uncertainty budget. The resulting uncertainty of the generated fields at the 1 sigma level is within 1.3*10/sup -3/ for the electric field up to 60 kV/m and within 6*10/sup -3/ for the magnetic field up to 10 mT.< >