Loading…

Development and Experimental Study of a Polytetrafluoroethylene-Tipped Microcatheter Poorly Adhesive to n -Butyl-2-Cyanoacrylate

PURPOSEWe have developed a new microcatheter (designated "NSX") with an outer layer of polytetrafluoroethylene (PTFE) at its tip. We compared the adhesion of the new NSX microcatheter and a conventional microcatheter with n-butyl-2-cyanoacrylate (NBCA) in vitro and in swine blood vessels.M...

Full description

Saved in:
Bibliographic Details
Published in:Journal of endovascular therapy 2023-10, p.15266028231208652-15266028231208652
Main Authors: Nakai, Motoki, Saguchi, Toru, Yunaiyama, Daisuke, Takara, Yuki, Tanaka, Taro, Okada, Yukinori, Saito, Kazuhiro
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:PURPOSEWe have developed a new microcatheter (designated "NSX") with an outer layer of polytetrafluoroethylene (PTFE) at its tip. We compared the adhesion of the new NSX microcatheter and a conventional microcatheter with n-butyl-2-cyanoacrylate (NBCA) in vitro and in swine blood vessels.MATERIALS AND METHODSThe 3 cm tip of the NSX microcatheter is composed of PTFE, which can be identified by double platinum markers. The tips of the NSX and conventional microcatheters were inserted into a vascular model filled with porcine blood with no flow, and NBCA mixed with lipiodol (1:2) was injected from the microcatheters. Two minutes after the injection of NBCA, the microcatheter was withdrawn and the degree of its adhesion to NBCA was evaluated by measuring the resistance value (N) during catheter removal with a digital force gauge. These measurements were repeated with 20 catheters of each type. Similarly, 5 injections were performed with both the NSX and conventional microcatheters in swine vessels. The degree of adhesion of the catheter and blood vessel was evaluated by 2 radiologists under X-ray fluoroscopy on a 3-point scale: 1, no adhesion; 2, mild adhesion; 3, strong adhesion.RESULTSThe mean resistance values (N) for the NSX and conventional microcatheters were 0.503±0.186 and 1.051±0.367 (N), respectively (p
ISSN:1526-6028
1545-1550
DOI:10.1177/15266028231208652