Biomechanical and histomorphometric study on the bone–screw interface of bioactive ceramic-coated titanium screws

The purpose of this study was to compare the osseointegration of 4 different kinds of bioactive ceramic-coated screws with uncoated screws by biomechanical and histomorphometric analysis. Calcium pyrophosphate (CPP), apatite-wollastonite 1:3 glass ceramic (W3G), apatite-wollastonite 1:1 glass cerami...

Full description

Saved in:
Bibliographic Details
Published in:Biomaterials 2005-06, Vol.26 (16), p.3249-3257
Main Authors: Hyup Lee, Jae, Ryu, Hyun-Seung, Lee, Dong-Soo, Sun Hong, Kug, Chang, Bong-Soon, Lee, Choon-Ki
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The purpose of this study was to compare the osseointegration of 4 different kinds of bioactive ceramic-coated screws with uncoated screws by biomechanical and histomorphometric analysis. Calcium pyrophosphate (CPP), apatite-wollastonite 1:3 glass ceramic (W3G), apatite-wollastonite 1:1 glass ceramic (WAG) and bioactive CaO–SiO 2–B 2O 3 glass ceramic (CSG) coatings were prepared and coated by the dipping method. Coated and uncoated titanium screws were inserted into the tibia of 18 adult mongrel male dogs for 2, 4, and 8 weeks. The insertion torques, radiographs, undecalcified histology, histomorphometric analysis, and extraction torques were evaluated. No difference of insertion torque was found among the five screw types. At 2 and 4 weeks after implantation, the extraction torque of ceramic-coated screws was significantly higher than that of uncoated screws ( p⩽0.0001). At 8 weeks, the extraction torques of CPP-, W3G-, and WAG-coated screws were significantly higher than those of CSG-coated and uncoated screws ( p
ISSN:0142-9612
1878-5905
DOI:10.1016/j.biomaterials.2004.08.033