Use of a modified microvascular plug to modify fenestration flow in a rapidly failing Fontan
We report a case of an alternative transcatheter use of the modified Medtronic microvascular plug to modify fenestration stent flow in a patient with a rapidly deteriorating clinical condition. This four-year-old boy developed severe cyanosis following fenestration stent insertion, initially placed...
Saved in:
Published in: | Cardiology in the young 2024-03, Vol.34 (3), p.698-700 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We report a case of an alternative transcatheter use of the modified Medtronic microvascular plug to modify fenestration stent flow in a patient with a rapidly deteriorating clinical condition. This four-year-old boy developed severe cyanosis following fenestration stent insertion, initially placed due to prolonged drainage post-Fontan with extra-cardiac conduit. In April 2023, he underwent urgent cardiac catheterisation and had partial occlusion of fenestration stent with a modified 9Q microvascular plug. His oxygen saturations improved from 50 to 89% in room air with no re-emergence of raised cavopulmonary pressures. |
---|---|