Loading…

Transmission grating spectrograph with on-line recording of XUV soft X-ray spectra from laser produced plasmas

The setting up and on-line operation of a transmission grating spectrograph in the XUV soft X-ray region is described. The detector was a microchannel plate–image intensifier combination, and the output of the image intensifier was coupled to a CCD camera–frame grabber system through an imaging lens...

Full description

Saved in:
Bibliographic Details
Published in:Optics and laser technology 2006-02, Vol.38 (1), p.46-50
Main Authors: Sailaja, S., Arora, V., Kumbhare, S.R., Joshi, R.A., Naik, P.A., Gupta, P.D.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The setting up and on-line operation of a transmission grating spectrograph in the XUV soft X-ray region is described. The detector was a microchannel plate–image intensifier combination, and the output of the image intensifier was coupled to a CCD camera–frame grabber system through an imaging lens. The spectrograph could be operated in the 5–20 Å range with 0.6 Å spectral resolution and in the 5–50 Å range with ∼1 Å resolution, respectively. The high sensitivity of the detector enables single shot operation, which is useful for several laser plasma interaction studies.
ISSN:0030-3992
1879-2545
DOI:10.1016/j.optlastec.2004.10.004