Determination of thePrpand radial dose correction factor in reference dosimetry
Objectives.In an addendum to AAPM TG-51 protocol, McEwenet al, (DOI:10.1118/1.4866223) introduced a new factorPrpto account for the radial dose distribution of the photon beam over the detector volume mainly in flattening filter free (FFF) beams.Prpand its extension to non-FFF beam reference dosimet...
Saved in:
Published in: | Biomedical physics & engineering express 2024-02, Vol.10 (2) |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Objectives.In an addendum to AAPM TG-51 protocol, McEwenet al, (DOI:10.1118/1.4866223) introduced a new factorPrpto account for the radial dose distribution of the photon beam over the detector volume mainly in flattening filter free (FFF) beams.Prpand its extension to non-FFF beam reference dosimetry is investigated to see its impact in a clinical situation.Approches.ThePrpwas measured using simplified version of Sudhyadhomet al(DOI:10.1118/1.4941691) for Elekta and Varian FFF beams with two commonly used calibration detectors; PTW-30013 and Exradin-A12 ion chambers after acquiring high resolution profiles in detectors cardinal coordinates. For radial dose correction factor, the ion chambers were placed in a small water phantom and the central axis position was set to center of the sensitive volume on the treatment table and was studied by rotating the table by 15-degree interval from -90 to +90 degrees with respect to the initial (zero) position.Main results.The magnitude ofPrpvaries very little with machine, detector and beam energies to a value of 1.003 ± 0.0005 and 1.005 ± 0.0005 for 6FFF and 10FFF, respectively. The radial anisotropy for the Elekta machine with Exradin-A12 and PTW-30013 detector the magnitudes are in the range of (0.9995±0.0011 to 1.0015±0.0010) and (0.9998±0.0007 to 1.0015±0.0010), respectively. Similarly, for the Varian machine with Exradin-A12 and PTW-30013 ion chambers, the magnitudes are in the range of (1.0004±0.0010 to 1.0018±0.0018) and (1.0006±0.0009 to 1.0027±0.0007), respectively.Significance.ThePrpis ≤ 0.3% and 0.5% for 6FFF and 10FFF, respectively. The radial dose correction factor in regular beams also does not impact the dosimetry where the maximum magnitude is ±0.2% which is within experimental uncertainty.Objectives.In an addendum to AAPM TG-51 protocol, McEwenet al, (DOI:10.1118/1.4866223) introduced a new factorPrpto account for the radial dose distribution of the photon beam over the detector volume mainly in flattening filter free (FFF) beams.Prpand its extension to non-FFF beam reference dosimetry is investigated to see its impact in a clinical situation.Approches.ThePrpwas measured using simplified version of Sudhyadhomet al(DOI:10.1118/1.4941691) for Elekta and Varian FFF beams with two commonly used calibration detectors; PTW-30013 and Exradin-A12 ion chambers after acquiring high resolution profiles in detectors cardinal coordinates. For radial dose correction factor, the ion chambers were placed in a small |
---|---|
ISSN: | 2057-1976 2057-1976 |
DOI: | 10.1088/2057-1976/ad25bc |